Z-2080-2019 Class II

Recalled by Bard Peripheral Vascular Inc — Tempe, AZ

Recall Details

Product Type
Devices
Report Date
July 31, 2019
Initiation Date
June 25, 2018
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
95 units

Product Description

POWERPORT CLEARVUE isp, 6F CHRONOFLEX Catheter, Intermediate NAUTILUS DELTA Kit; POWERPORT CLEARVUE isp, 8F CHRONOFLEX Catheter, Intermediate NAUTILUS DELTA Kit; POWERPORT CLEARVUE Slim, 6F CHRONOFLEX Catheter and Suture Plugs, Intermediate NAUTILUS DELTA Kit; POWERPORT CLEARVUE Slim, 8F CHRONOFLEX Catheter and Suture Plugs, Intermediate NAUTILUS DELTA Kit; NAUTILUS DELTA Tip Confirmation System (Includes: Netbook with Pre-loaded Software, Patient Module, and ECG Cable) Product Usage: Is an implantable access device designed to provide repeated access to the vascular system.

Reason for Recall

Certain NAUTILUS DELTA Port Kit product code / lot number combinations may be at risk of experiencing a lack of or a loss of the intravascular ECG signal during port placement procedures when used in combination with the NAUTILUS DELTA" TCS.

Distribution Pattern

US Nationwide Distribution in the states of FL, CO, WI, OH, IL, KS, AZ, WV, CT, PA, AL, MI and VA

Code Information

REBR1147, REBR1148, REBR1149, REBT0471, RECQ2162 and REBR1150