Z-0467-2025 Class II Ongoing

Recalled by Bard Peripheral Vascular Inc — Tempe, AZ

FDA device recall Z-0467-2025 was initiated by Bard Peripheral Vascular Inc on Sep 30, 2024 and is designated Class II. Reason for recall: Disposable biopsy instruments were potentially manufactured missing the backend notch retention feature on the needle that results in inadequate securing of the needle inside the device, which could… The recall status is ongoing. Affected quantity: 71,498.

Recall details

Product Type
Devices
Report Date
Nov 27, 2024
Initiation Date
Sep 30, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
71,498

Product description

Bard Marquee Disposable Core Biopsy Instrument Kit: BARD MARQUEE KIT 12G X 10 CM, REF: MQK1210; BARD MARQUEE 12G X 10 CM, REF: MQ1210; BARD MARQUEE Kit 12G X 13 CM, REF: MQK1213; BARD MARQUEE 12G X 13 CM, REF: MQ1213

Reason for recall

Disposable biopsy instruments were potentially manufactured missing the backend notch retention feature on the needle that results in inadequate securing of the needle inside the device, which could lead to needle detachment and misplacement, which could lead to tissue damage and bleeding.

Distribution pattern

Worldwide - US Nationwide including in the states of WA, LA, MN, PA, UT, SC, OK, KY, AZ, IL, TX, NJ, NY, FL, TN, IN, OH, NC, CO, VA, GA, MD, CA, AL, WI, MO, IA, CT, NE, MI, MS, NV, KS, OR, SD, ID, NH, RI, WY, AR, MA, DE, ND, NM, AK, PR and the countries of Portugal, Spain, South Africa, Greece, United Arab Emirates, United Kingdom, Sweden, Finland, Belgium, Korea, Australia, Canada.

Code information

REF/UDI-DI/Lot(Expiration): MQK1210/00801741097263/1490758(9/28/2025), 1506044(1/28/2026), 1506045(1/28/2026), 1511827(2/28/2026), 1514692(3/28/2026), 1518977(4/28/2026), 1518978(4/28/2026), 1518979(4/28/2026), 1519143(4/28/2026), 1520507(5/28/2026), 1520508(5/28/2026), 1522157(5/28/2026), 1522158(5/28/2026), 1522577(5/28/2026), 1523785(6/28/2026), 1523914(6/28/2026), 1523915(6/28/2026), 1525370(6/28/2026), 1525371(6/28/2026), 1525373(6/28/2026), 1528417(7/28/2026), 1528418(7/28/2026), 1528495(7/28/2026), 1528496(7/28/2026), 1528497(7/28/2026), 1528560(7/28/2026), 0001618867(3/28/2028), 0001620281(3/28/2028), 0001632675(6/28/2028); MQ1210/00801741097119/1514460(3/28/2026), 1514461(3/28/2026), 1519144(4/28/2026), 1519145(4/28/2026), 1519456(4/28/2026), 1520066(5/28/2026), 1520067(5/28/2026), 1520506(5/28/2026), 1523783(6/28/2026), 1523784(6/28/2026), 0001618258(3/28/2028), 0001631611(6/28/2028), 0001633216(7/28/2028); MQK1213/00801741097270/1543517(10/28/2026), 1559312(2/28/2027), 1562724(3/28/2027), 1569737(4/28/2027), 1581652(7/28/2027); MQ1213/00801741097126/1558191(2/28/2027), 1559313(2/28/2027)

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0467-2025
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Bard Peripheral Vascular Inc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0467-2025. Report a data issue: hello@regdatalab.com.