Z-0887-2026 Class II Ongoing

Recalled by Bard Peripheral Vascular Inc — Tempe, AZ

Recall Details

Product Type
Devices
Report Date
December 17, 2025
Initiation Date
November 6, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1240

Product Description

Safe-T-Centesis 6 Fr Catheter Drainage Tray, REF: PIG1260T,

Reason for Recall

Acute drainage catheters, indicated for use in removing fluid during acute drainage/aspiration procedure, manufactured with safety indicators that erroneously show catheter in sharp position when catheter is in blunt position, which may lead users to believe device is not yet in pleural or peritoneal space when it is, may cause discomfort, delay to treatment, injury to internal organs or vessels.

Distribution Pattern

US Nationwide distribution in the states of CA, MN, IN, MI, LA.

Code Information

UDI/Lot/Expiration: (01)10885403108365(17)271130(10)0001623320/0001623320/20271130; (01)10885403108365(17)271130 (10)0001623069/0001623069/20271130