Z-1244-2025 Class I Ongoing

Recalled by Bard Peripheral Vascular Inc — Tempe, AZ

FDA device recall Z-1244-2025 was initiated by Bard Peripheral Vascular Inc on Feb 5, 2025 and is designated Class I. Reason for recall: Atherectomy Catheter eIFU updated to clarify and emphasize procedural steps intended to reduce the likelihood of catheter breakage events. Catheter has an outer cylinder connected to a rotating helix… The recall status is ongoing. Affected quantity: 15,755.

Recall details

Product Type
Devices
Report Date
Mar 12, 2025
Initiation Date
Feb 5, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
15,755

Product description

Rotarex Atherectomy System and Instructions for Use, BD SET Rotarex: S 6 F x 110 cm, REF: 80236; S 6 F x 135 cm, REF: 80237; S 8 F x 85 cm, REF: 80238; S 8 F x 110 cm, REF: 80239

Reason for recall

Atherectomy Catheter eIFU updated to clarify and emphasize procedural steps intended to reduce the likelihood of catheter breakage events. Catheter has an outer cylinder connected to a rotating helix, which could fracture and/or break, which would require retrieval, and helix facture/break could cause vessel injury and lead to severe bleeding.

Distribution pattern

US Nationwide Distribution: NY, VA, KY, IN, OR, FL, MD, NJ, LA, CO, PR, PA, MS, MI, TX, KS, IL, CA, GA, OH, AL, NC, ND, AZ, SC, TN, UT, AR, WI, SD, MA, MO, ME, DE, WV , NV, NH, MN, CT, NE, IA, NM, WA, HI, and OK

Code information

Instructions for Use versions prior to ZE10895 revision C1 09/24. REF/UDI-DI/Lot(Expiration): 80236/07640142811855/230893(2025-12-15), 231100(2026-06-09), 231172(2026-06-09), 231171(2026-06-09), 231526(2026-08-21), 231875(2026-08-21), 240018(2026-11-22), 240017(2026-11-22), 240375(2027-01-18), 240567(2027-01-18), 241075(2024-05-22), 241335(2024-06-19), 241550(2024-07-19), 241777(2024-08-19), 242691(2024-12-13); 80237/07640142811862/230897(2026-04-11), 230898(2026-04-11), 231033(2026-05-31), 231101(2026-06-13), 231170(2026-06-13), 231229(2026-07-07), 231400(2026-07-14), 231414(2026-07-31), 231412(2026-06-30), 231413(2026-07-31), 231411(2026-08-03), 231521(2026-09-01), 231520(2026-07-31), 231522(2026-08-31), 231523(2026-09-01), 231524(2026-09-01), 231525(2026-09-01), 231715(2026-09-22), 231885(2026-09-22), 231884(2026-10-27), 240020(2026-11-22), 240396(2026-12-08), 240530(2027-01-10), 240531(2027-01-10), 240566(2027-01-10), 240568(2027-01-25), 240754(2027-02-06), 240769(2027-02-09), 240795(2027-03-04), 241074(2027-03-07), 241119(2027-03-21), 241190(2027-04-16), 241325(2027-04-16), 241328(2027-04-25), 241336(2027-04-29), 241464(2027-05-07), 241471(2027-05-27), 241553(2027-06-05), 242019(2027-06-28), 242023(2027-06-12), 242057(2027-06-20), 242056(2027-06-20), 242112(2027-06-26), 242160(2027-07-25), 242321(2027-07-08), 242493(2027-08-01), 242596(2027-09-20), 242693(2027-11-06); 80238/07640142811879/230895(2025-11-23), 240128(2026-05-31), 240276(2026-05-31), 240532(2026-05-31), 240658(2026-09-30), 241076(2026-09-30), 241330(2026-09-30), 241548(2026-09-30), 241768(2026-09-30), 241911(2026-09-30), 242117(2026-09-30), 242494(2026-09-30), 242521(2026-09-30), 242519(2026-05-31), 242520(2026-09-30); 80239/07640142811886/230896(2025-11-23), 231231(2026-06-13), 231410(2026-06-13), 231527(2026-08-31), 231883(2026-10-19), 232029(2026-08-31), 232028(2026-10-19), 240129(2026-11-30), 240278(2026-11-30), 240557(2027-01-22), 240746(2027-01-22), 240794(2027-02-28), 241077(2027-04-08), 241188(2027-04-08), 241462(2027-05-03), 241523(2027-05-03), 241767(2027-06-26), 241910(2027-06-26), 242159(2027-08-09), 242322(2027-09-10), 242518(2027-09-10), 242694(2027-10-25), 242695(2027-10-25)

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1244-2025
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Bard Peripheral Vascular Inc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1244-2025. Report a data issue: hello@regdatalab.com.