Z-2082-2016 Class II Terminated
FDA device recall Z-2082-2016 was initiated by Medtronic Navigation, Inc.-Littleton on Mar 17, 2016 and is designated Class II. Reason for recall: It has been discovered that replacement x-ray generator and system motion batteries could become damaged during shipment, resulting in unacceptable risk of failure. The recall status is terminated (terminated Apr 6, 2021). Affected quantity: US - 581.
Recall details
- Product Type
- Devices
- Report Date
- Jul 13, 2016
- Initiation Date
- Mar 17, 2016
- Termination Date
- Apr 6, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- US - 581
Product description
Medtronic 0-arm Imaging mobile x-ray system. Model Numbers - B1-700-00027, B1-700-00028
Reason for recall
It has been discovered that replacement x-ray generator and system motion batteries could become damaged during shipment, resulting in unacceptable risk of failure.
Distribution pattern
Nationwide Distribution
Code information
Model No. BI-700-00027 and BT-700-00028
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2082-2016
- openFDA file date
- Nov 17, 2026
- Loaded by RegDataLab