Z-2083-2015 Class II Terminated
Recalled by Maquet Cardiovascular Us Sales, Llc
— Wayne, NJ
FDA device recall Z-2083-2015 was initiated by Maquet Cardiovascular Us Sales, Llc on Mar 2, 2015 and is designated Class II. Reason for recall: Maquet has identified that in some FLOW-i Anesthesia System Auxiliary O2 and Suction Modules, the suctioning function cannot be activated or controlled by the suction unit on/off switch leading to a… The recall status is terminated (terminated Apr 29, 2016). Affected quantity: 104 units.
Recall details
- Product Type
- Devices
- Report Date
- Jul 22, 2015
- Initiation Date
- Mar 2, 2015
- Termination Date
- Apr 29, 2016
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 104 units
Product description
The FLOW-i Anesthesia System Axillary O2 and Suction Module, Device Part number: 66 79 847 FLOW-i Anesthesia System: Is indicated for administering inhalation anesthesia while controlling the entire ventilation of patients with no ability to breathe, as well as in supporting patients with a limited ability to breathe.The system is intended for use in a wide range of population: from neonatal to adult patients. The system is intended for use in hospital environments, except for the MRI environment, by healthcare professionals trained in inhalation anesthesia administration. Auxiliary O2 and Suction Module: Is an option available for the FLOW-i Anesthesia System. The module contains two main parts: Flowmeter unit: The intended use of the flowmeter is to provide oxygen for patient therapy. Suction unit: The intended use is to extract body fluids from the stomach and airways.
Reason for recall
Maquet has identified that in some FLOW-i Anesthesia System Auxiliary O2 and Suction Modules, the suctioning function cannot be activated or controlled by the suction unit on/off switch leading to a decrease or loss of suctioning capability of the modules.
Distribution pattern
Worldwide Distribution - US (nationwide) and Internationally to Australia, Belgium, Brazil, China, Colombia, Croatia, Czech Republic, Ecuador, Finland, France, Germany, Hong Kong, India, Israel, Japan, Mexico, Netherlands, Peru, Poland, Portugal, Romania, Russia, Serbia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand and United Arab Emirates.
Code information
Device Part number: 66 79 847 Serial Numbers: 4, 14-24, 26-29, 31, 37, 40, 56, 138, 199, 205, 367, 369-372, 391, 394-395, 416, 466-491, 517, 519, 525, 567-568, 570, 582-583, 606, 613, 619, 627, 629-631, 633, 635-641, and 643-665
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2083-2015
- openFDA file date
- Nov 16, 2026
- Loaded by RegDataLab
- Nov 17, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.