Z-2084-2015 Class II Terminated

Recalled by Northern Digital Inc. — Waterloo, Ontario

Recall Details

Product Type
Devices
Report Date
July 22, 2015
Initiation Date
June 10, 2015
Termination Date
October 3, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
524, 700 spheres (41773 = 440,820 spheres; 41774 = 77,760 spheres and 41813 = 6,120 spheres)

Product Description

NDI Disposable Reflective Marker Spheres for Brainlab IGS (Image Guided Surgical) system. Product Usage: Disposable reflective marker spheres are attached to reference arrays and instruments, thus enabling the infrared tracking systems to detect the position of the patient and instruments in the surgical field.

Reason for Recall

Inadequate curing of adhesive, may cause the two halves of the spheres to separate.

Distribution Pattern

US Nationwide Distribution

Code Information

1) Catalog Number: 41773 (90 pack) All Lots beginning with the prefix C / All expiration dates; 2) Catalog Number: 41774 (270 pack) All Lots beginning with the prefix C / All expiration dates; and 3) Catalog Number: 41813 (60 pack) All Lots beginning with the prefix C / All expiration dates