FDA device recall Z-2084-2016 was initiated by Integra LifeSciences Corp. on May 20, 2016 and is designated Class II. Reason for recall: The affected K-wires do not meet specification requirements. Specifically, a) K-wires were made of 316LVM stainless steel instead of 304 stainless steel. b) K-wires have passivated surface finish… The recall status is terminated (terminated Nov 16, 2016). Affected quantity: 182 units.
Recall details
- Product Type
- Devices
- Report Date
- Jul 6, 2016
- Initiation Date
- May 20, 2016
- Termination Date
- Nov 16, 2016
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 182 units
Product description
K-wire, Smooth Double-pointed 0.8 x 90mm, a component of the METASURG Capture Screw System The K-wire is used in conjunction with the 2.0 mm cannulated screws for the METASURG Capture Screw System. The K-wire is used by the physician to insert into the bone and to act as a guide for drilling, countersinking and installing screws
Reason for recall
The affected K-wires do not meet specification requirements. Specifically, a) K-wires were made of 316LVM stainless steel instead of 304 stainless steel. b) K-wires have passivated surface finish instead of electro-polish surface finish. c) K-wires have incorrect length and diameter.
Distribution pattern
Nationwide Distribution to OK, CA, TX, WA, GA, NY, MS, NY, PA
Code information
Lot 237830TE
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2084-2016
- openFDA file date
- Nov 16, 2026
- Loaded by RegDataLab
- Nov 17, 2026