Z-2090-2012 Class II Ongoing
FDA device recall Z-2090-2012 was initiated by Stryker Howmedica Osteonics Corp. on Apr 30, 2012 and is designated Class II. Reason for recall: Stryker has updated the instruction for use (IFU) (product correction)for the ABGII Modular and Rejuvenate Modular Hip Systems. This is based on a reported rate of less than one percent for revisions… The recall status is ongoing. Affected quantity: (NLS/NLV) - 23193 units and (SPT) - 20,743 units.
Recall details
- Product Type
- Devices
- Report Date
- Aug 1, 2012
- Initiation Date
- Apr 30, 2012
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- (NLS/NLV) - 23193 units and (SPT) - 20,743 units
Product description
Rejuvenate Modular Stems Stryker Ireland Carrigtwohill Industrial Estate Carrigtwohill County Cork, Ireland Rejuvenate Modular Necks Stryker Howmedica Osteonics Corp. Mahwah, NJ 07430. Stem Bodies Rejuvenate Modular consists of 6 modular stems bodies in sizes 7-12. The Modular Stem bodies are made of TMZF alloy, developed specifically for orthopaedic applications. Rejuvenate Modular Necks The Rejuvenate Modular Necks are made of Cobalt Chrome. The necks in this system consist of 4 lengths in 4 mm increments. They are 30 mm, 34 mm, 38 mm and 42 mm. They come in multiple neck angles and versions in each length. They are 127, 130 and 132. The versions are 0, 8, 16 degrees. In total there are 16 unique neck implants in the Rejuvenate Modular System.
Reason for recall
Stryker has updated the instruction for use (IFU) (product correction)for the ABGII Modular and Rejuvenate Modular Hip Systems. This is based on a reported rate of less than one percent for revisions potentially associated with fretting and or corrosion at or about the modular neck junction. On June 29, 2012, Stryker has decided to remove the affected product from the field since there may be an
Distribution pattern
Nationwide Distribution
Code information
510 K092561 all lots Rejuvenate (NLS/NLV) Rejuvenate (SPT) (4845-4XXX and NLS/NLV)
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2090-2012
- openFDA file date
- Nov 16, 2026
- Loaded by RegDataLab