Z-2090-2012 Class II Ongoing

Recalled by Stryker Howmedica Osteonics Corp. — Mahwah, NJ

FDA device recall Z-2090-2012 was initiated by Stryker Howmedica Osteonics Corp. on Apr 30, 2012 and is designated Class II. Reason for recall: Stryker has updated the instruction for use (IFU) (product correction)for the ABGII Modular and Rejuvenate Modular Hip Systems. This is based on a reported rate of less than one percent for revisions… The recall status is ongoing. Affected quantity: (NLS/NLV) - 23193 units and (SPT) - 20,743 units.

Recall details

Product Type
Devices
Report Date
Aug 1, 2012
Initiation Date
Apr 30, 2012
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
(NLS/NLV) - 23193 units and (SPT) - 20,743 units

Product description

Rejuvenate Modular Stems Stryker Ireland Carrigtwohill Industrial Estate Carrigtwohill County Cork, Ireland Rejuvenate Modular Necks Stryker Howmedica Osteonics Corp. Mahwah, NJ 07430. Stem Bodies Rejuvenate Modular consists of 6 modular stems bodies in sizes 7-12. The Modular Stem bodies are made of TMZF alloy, developed specifically for orthopaedic applications. Rejuvenate Modular Necks The Rejuvenate Modular Necks are made of Cobalt Chrome. The necks in this system consist of 4 lengths in 4 mm increments. They are 30 mm, 34 mm, 38 mm and 42 mm. They come in multiple neck angles and versions in each length. They are 127, 130 and 132. The versions are 0, 8, 16 degrees. In total there are 16 unique neck implants in the Rejuvenate Modular System.

Reason for recall

Stryker has updated the instruction for use (IFU) (product correction)for the ABGII Modular and Rejuvenate Modular Hip Systems. This is based on a reported rate of less than one percent for revisions potentially associated with fretting and or corrosion at or about the modular neck junction. On June 29, 2012, Stryker has decided to remove the affected product from the field since there may be an

Distribution pattern

Nationwide Distribution

Code information

510 K092561 all lots Rejuvenate (NLS/NLV) Rejuvenate (SPT) (4845-4XXX and NLS/NLV)

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2090-2012
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2090-2012. Report a data issue: hello@regdatalab.com.