Z-2090-2026 Class II Ongoing

Recalled by Medline Industries, LP — Northfield, IL

FDA device recall Z-2090-2026 was initiated by Medline Industries, LP on January 7, 2026 and is designated Class II. Reason for recall: Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however,… The recall status is ongoing. Affected quantity: 1774 units.

Recall Details

Product Type
Devices
Report Date
May 13, 2026
Initiation Date
January 7, 2026
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1774 units

Product Description

Medline Convenience Kits: 1) SCC RF PAIN MGMT, Model Number:DYNJ56436A; 2) SCC RF PAIN MGMT, Model Number:DYNJ56436B; 3) SCC RF PAIN MGMT, Model Number:DYNJ56436C; 4) CPNB SETUP TRAY, Model Number:DYNJRA1979; 5) CPNB SETUP TRAY, Model Number:DYNJRA1979A; 6) CPNB SETUP TRAY, Model Number:DYNJRA1979B; 7) VARITHENA TRAY 2, Model Number:DYNJRA2608; 8) CPNB SETUP TRAY, Model Number:SAMPA0108

Reason for Recall

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

Distribution Pattern

Worldwide - US Nationwide distribution.

Code Information

1) DYNJ56436A, UDI-DI: 10889942749874(each), 40889942749875(case), Lot Number: 21AMB799; 2) DYNJ56436A, UDI-DI: 10889942749874(each), 40889942749875(case), Lot Number: 21BMC737; 3) DYNJ56436B, UDI-DI: 10193489869569(each), 40193489869560(case), Lot Number: 21DME244; 4) DYNJ56436C, UDI-DI: 10195327059668(each), 40195327059669(case), Lot Number: 23AMG574; 5) DYNJ56436C, UDI-DI: 10195327059668(each), 40195327059669(case), Lot Number: 23DMA829; 6) DYNJ56436C, UDI-DI: 10195327059668(each), 40195327059669(case), Lot Number: 23DMI776; 7) DYNJ56436C, UDI-DI: 10195327059668(each), 40195327059669(case), Lot Number: 23EMH095; 8) DYNJ56436C, UDI-DI: 10195327059668(each), 40195327059669(case), Lot Number: 23FMI704; 9) DYNJ56436C, UDI-DI: 10195327059668(each), 40195327059669(case), Lot Number: 23HMC628; 10) DYNJ56436C, UDI-DI: 10195327059668(each), 40195327059669(case), Lot Number: 23IMC652; 11) DYNJRA1979, UDI-DI: 10195327186852(each), 40195327186853(case), Lot Number: 22GBV936; 12) DYNJRA1979, UDI-DI: 10195327186852(each), 40195327186853(case), Lot Number: 22IBU628; 13) DYNJRA1979A, UDI-DI: 10195327649258(each), 40195327649259(case), Lot Number: 24IMG516; 14) DYNJRA1979A, UDI-DI: 10195327649258(each), 40195327649259(case), Lot Number: 24JMA460; 15) DYNJRA1979A, UDI-DI: 10195327649258(each), 40195327649259(case), Lot Number: 24KMA261; 16) DYNJRA1979A, UDI-DI: 10195327649258(each), 40195327649259(case), Lot Number: 25DME153; 17) DYNJRA1979B, UDI-DI: 10198459371950(each), 40198459371951(case), Lot Number: 25FBG333; 18) DYNJRA2608, UDI-DI: 10198459331640(each), 40198459331641(case), Lot Number: 25ELB099; 19) SAMPA0108, UDI-DI: 10195327096861(each), 40195327096862(case), Lot Number: 22BBD381; 20) SAMPA0108, UDI-DI: 10195327096861(each), 40195327096862(case), Lot Number: 22NBG086