Z-2097-2015 Class II Terminated

Recalled by Maquet Cardiovascular Us Sales, Llc — Wayne, NJ

FDA device recall Z-2097-2015 was initiated by Maquet Cardiovascular Us Sales, Llc on Jun 10, 2015 and is designated Class II. Reason for recall: There is a potential issue that could result to collision between the C-arm of the angiography system and the MAGNUS OR table system. After switching on the angiography-system, in rare cases, MAGNUS… The recall status is terminated (terminated Apr 18, 2016). Affected quantity: 12 units.

Recall details

Product Type
Devices
Report Date
Jul 29, 2015
Initiation Date
Jun 10, 2015
Termination Date
Apr 18, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
12 units

Product description

Magnus Hybrid OR Table Column (1180.01B2) used in conjunction with Siemens Artis Zeego Angiography System. Maquet GmbH & Co.

Reason for recall

There is a potential issue that could result to collision between the C-arm of the angiography system and the MAGNUS OR table system. After switching on the angiography-system, in rare cases, MAGNUS column provides wrong position data for the longitudinal movement leading to a collision.

Distribution pattern

Worldwide Distribution - US Nationwide Distribution and to the countries of : France, Australia, Belgium, Brazil, Switzerland, Chile, China, Germany, Spain, United Kingdom, Hong Kong, Ireland, Israel, India, Italy, Japan, Kuwait, Kazakhstan, Mexico, Netherlands, Norway, Russia, Sweden, Turkey, Taiwan and South Africa.

Code information

SN 00048

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2097-2015
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2097-2015. Report a data issue: hello@regdatalab.com.