Z-2099-2015 Class II Terminated

Recalled by Stryker Instruments Div. Of Stryker Corporation — Portage, MI

FDA device recall Z-2099-2015 was initiated by Stryker Instruments Div. Of Stryker Corporation on May 22, 2015 and is designated Class II. Reason for recall: Stryker Instruments is initiating a recall of the Neptune 2 Waste Management System due to complaints related to installation failures of the Neptune 2 diverter valve assembly during docking. The recall status is terminated (terminated Sep 12, 2016). Affected quantity: 9,245.

Recall details

Product Type
Devices
Report Date
Jul 29, 2015
Initiation Date
May 22, 2015
Termination Date
Sep 12, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
9,245

Product description

0702-001-350Q, Diverter Valve Assembly 0702-001-000, Neptune 2 Rover Ultra (120V) 0702-001-000R, Repair Neptune 2 Rover 0702-002-000, Neptune 2 Rover Ultra (230V) 0702-002-000ES, Neptune 2 Rover (230V) Spanish 0702-002-000FI, Neptune 2 Ultra Rover (230V) Finnish 0702-002-000IT, Neptune 2 Rover (230V) Italian 0702-002-000R, Repair Neptune 2 Rover Ultra (230V) Product Usage: The Neptune 2 Waste Management System is intended to be used in the operating room, pathology, surgical centers, and doctor s offices to collect and dispose of surgical fluid waste as well as collect smoke generated from electro cautery or laser devices. Contraindication: connection of the Neptune 2 Waste Management System directly to chest tubes or other closed wound drains. The Stryker Neptune 2 Rover is a mobile unit used to suction, collect fluid waste, and evacuate surgical smoke from a surgical site during operation. Fluid waste is removed from the surgical site through suction tubing connected to inlet ports of disposable manifolds installed in the rover. The Diverter Valve Assembly consists of two diverter valves with three hoses, one inlet and two outlets. The sole purpose of the Diverter Valve Assembly is to direct/divert water from the fresh water inlet to either the large canister or small canister via the pre-fill tank.

Reason for recall

Stryker Instruments is initiating a recall of the Neptune 2 Waste Management System due to complaints related to installation failures of the Neptune 2 diverter valve assembly during docking.

Distribution pattern

Worldwide Distribution in the states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN. KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, ND, NE, NH, NY, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and the countries of: Canada, Europe, Asia Pacific, EEMEA.

Code information

Serial Numbers ranging from 0803118093 through 1505400223; Distribution or Repair dates: November 1, 2011 to March 31, 2015

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2099-2015
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2099-2015. Report a data issue: hello@regdatalab.com.