Z-2099-2017 Class I Terminated

Recalled by Cook Medical Incorporated — Bloomington, IN

FDA device recall Z-2099-2017 was initiated by Cook Medical Incorporated on Mar 22, 2017 and is designated Class I. Reason for recall: There have been five reports of graft thrombosis/occlusion during global commercial use, each following treatment for BTAI. One case resulted in patient death, and three cases resulted in… The recall status is terminated (terminated Feb 14, 2019). Affected quantity: 4916.

Recall details

Product Type
Devices
Report Date
Aug 16, 2017
Initiation Date
Mar 22, 2017
Termination Date
Feb 14, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4916

Product description

Zenith Alpha Thoracic Endovascular Graft The Zenith Alpha" Thoracic Endovascular Graft is indicated for the endovascular treatment of patients with isolated lesions of the descending thoracic aorta (not including dissections) having vascular anatomy suitable for endovascular repair including: " Iliac/femoral anatomy that is suitable for access with the required introduction systems, " Nonaneurysmal aortic segments (fixation sites) proximal and distal to the thoracic lesion: " with a length of at least 20 mm, and " with a diameter measured outer wall to outer wall of no greater than 42 mm and no less than 15 mm.

Reason for recall

There have been five reports of graft thrombosis/occlusion during global commercial use, each following treatment for BTAI. One case resulted in patient death, and three cases resulted in reintervention.

Distribution pattern

Nationwide Distribution.

Code information

All lots. Catalog Numbers: ZTA-D-28-160-W ZTA-D-28-229-W ZTA-D-30-160-W ZTA-D-30-229-W ZTA-D-32-160-W ZTA-D-32-229-W ZTA-D-34-142-W ZTA-D-34-190-W ZTA-D-36-142-W ZTA-D-36-190-W ZTA-D-38-147-W ZTA-D-38-197-W ZTA-D-40-147-W ZTA-D-40-197-W ZTA-D-42-152-W ZTA-D-42-204-W ZTA-D-44-157-W ZTA-D-44-211-W ZTA-D-46-157-W ZTA-D-46-211-W ZTA-DE-18-104-W ZTA-DE-18-148-W ZTA-DE-20-104-W ZTA-DE-20-148-W ZTA-DE-22-104-W ZTA-DE-22-148-W ZTA-DE-24-104-W ZTA-DE-24-148-W ZTA-DE-26-104-W ZTA-DE-26-148-W ZTA-DE-28-108-W ZTA-DE-28-154-W ZTA-DE-30-108-W ZTA-DE-30-154-W ZTA-DE-32-108-W ZTA-DE-32-154-W ZTA-DE-34-112-W ZTA-DE-34-160-W ZTA-DE-36-112-W ZTA-DE-36-160-W ZTA-DE-38-141-W ZTA-DE-38-91-W ZTA-DE-40-141-W ZTA-DE-40-91-W ZTA-DE-42-146-W ZTA-DE-42-94-W ZTA-DE-44-151-W ZTA-DE-44-97-W ZTA-DE-46-151-W ZTA-DE-46-97-W ZTA-P-18-105-W ZTA-P-18-127-W ZTA-P-20-105-W ZTA-P-20-127-W ZTA-P-22-105-W ZTA-P-22-127-W ZTA-P-24-105-W ZTA-P-24-127-W ZTA-P-26-105-W ZTA-P-26-149-W ZTA-P-28-109-W ZTA-P-28-132-W ZTA-P-28-155-W ZTA-P-28-201-W ZTA-P-30-109-W ZTA-P-30-132-W ZTA-P-30-155-W ZTA-P-30-201-W ZTA-P-32-109-W ZTA-P-32-132-W ZTA-P-32-155-W ZTA-P-32-201-W ZTA-P-34-113-W ZTA-P-34-137-W ZTA-P-34-161-W ZTA-P-34-209-W ZTA-P-36-113-W ZTA-P-36-137-W ZTA-P-36-161-W ZTA-P-36-209-W ZTA-P-38-117-W ZTA-P-38-142-W ZTA-P-38-167-W ZTA-P-38-217-W ZTA-P-40-117-W ZTA-P-40-142-W ZTA-P-40-167-W ZTA-P-40-217-W ZTA-P-42-121-W ZTA-P-42-147-W ZTA-P-42-173-W ZTA-P-42-225-W ZTA-P-44-125-W ZTA-P-44-152-W ZTA-P-44-179-W ZTA-P-44-233-W ZTA-P-46-125-W ZTA-P-46-152-W ZTA-P-46-179-W ZTA-P-46-233-W ZTA-PT-22-18-105-W ZTA-PT-26-22-105-W ZTA-PT-30-26-108-W ZTA-PT-32-28-178-W ZTA-PT-32-28-201-W ZTA-PT-34-30-161-W ZTA-PT-34-30-209-W ZTA-PT-36-32-161-W ZTA-PT-36-32-209-W ZTA-PT-38-34-167-W ZTA-PT-38-34-217-W ZTA-PT-40-36-167-W ZTA-PT-40-36-217-W ZTA-PT-42-38-173-W ZTA-PT-42-38-225-W ZTA-PT-44-40-179-W ZTA-PT-44-40-233-W ZTA-PT-46-42-179-W ZTA-PT-46-42-233-W

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2099-2017
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Cook Medical Incorporated? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Cook Medical Incorporated: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Cook Medical Incorporated →

Source: openFDA record for Z-2099-2017. Report a data issue: hello@regdatalab.com.