Z-2106-2020 Class II Ongoing

Recalled by Medtronic Inc. — Northridge, CA

FDA device recall Z-2106-2020 was initiated by Medtronic Inc. on May 20, 2019 and is designated Class II. Reason for recall: Customers using the firm's continuous glucose monitoring system application on an iPhone, iPad or iPod Touch with ios software version 12, 12.1, or 12,2 are likely to experience a shortened… The recall status is ongoing. Affected quantity: 23,978.

Recall details

Product Type
Devices
Report Date
Jun 3, 2020
Initiation Date
May 20, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
23,978

Product description

Guardian Connect App CSS7200 iOS and Guardian Connect Transmitter GST4C used on the iPhone, iPad and iPod Touch devices.

Reason for recall

Customers using the firm's continuous glucose monitoring system application on an iPhone, iPad or iPod Touch with ios software version 12, 12.1, or 12,2 are likely to experience a shortened transmitter battery life (approximately 4-5 days instead of the normal 7 days or more) after a full charge.

Distribution pattern

US Nationwide distribution.

Code information

Model Number CSS7200 (Guardian Connect App: iOS); Catalog Numbers: MMT-7821 (Guardian Connect Transmitter - GST4C); UPN - 00763000186333. There are no assigned lot or serial numbers.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2106-2020
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Medtronic Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Medtronic Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Medtronic Inc. →

Source: openFDA record for Z-2106-2020. Report a data issue: hello@regdatalab.com.