Z-2107-2020 Class II Terminated

Recalled by Quest Medical, Inc. — Allen, TX

FDA device recall Z-2107-2020 was initiated by Quest Medical, Inc. on Dec 17, 2019 and is designated Class II. Reason for recall: Additive and Arrest Agent Labels on the MPS Top Cover Replacement Kit, a replacement for worn or damaged MPS Console Top Covers for cardiopulmonary bypass heat exchanger, used to deliver whole blood… The recall status is terminated (terminated Apr 17, 2024). Affected quantity: 5 units.

Recall details

Product Type
Devices
Report Date
Jun 3, 2020
Initiation Date
Dec 17, 2019
Termination Date
Apr 17, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5 units

Product description

Ref 801168, MPS Top Cover Replacement Kit, 1 UNITS, UDI: (01)2 0634624 81168 7 (11)190916 (17)290913 (10)059135 (The device is a component of the MPS Console used as replacements for worn or damaged MPS Console top covers)

Reason for recall

Additive and Arrest Agent Labels on the MPS Top Cover Replacement Kit, a replacement for worn or damaged MPS Console Top Covers for cardiopulmonary bypass heat exchanger, used to deliver whole blood to the heart during open heart surgery, were swapped and incorrectly placed. This could lead to over delivery of potassium by the user requiring medical intervention.

Distribution pattern

US Nationwide distribution including in the states of NY, PA, TN, WV. OUS: None

Code information

Lot Number: 059135

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2107-2020
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Quest Medical, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2107-2020. Report a data issue: hello@regdatalab.com.