Z-2107-2023 Class II Ongoing

Recalled by Osteomed, LLC — Addison, TX

FDA device recall Z-2107-2023 was initiated by Osteomed, LLC on May 31, 2023 and is designated Class II. Reason for recall: Screws for osteotomies have contain an undersize screwhead which has the potential to pass through the plate hole. The recall status is ongoing.

Recall details

Product Type
Devices
Report Date
Jul 19, 2023
Initiation Date
May 31, 2023
Voluntary/Mandated
Voluntary: Firm initiated

Product description

REF: 218-1604, 1.6mm x 4mm Profile Zero Auto-Drive Screw, Rx Only, Sterile

Reason for recall

Screws for osteotomies have contain an undersize screwhead which has the potential to pass through the plate hole.

Distribution pattern

Worldwide - US Nationwide distribution in the states of AZ, CA, FL, GA, KS, KY, MD, MO, MT, NE, NY, TN, TX and the country of SG.

Code information

Lot # 1177027, 1177718, 1177812 / UDI-DI: (01)00845694034258