Z-2107-2023 Class II Ongoing
FDA device recall Z-2107-2023 was initiated by Osteomed, LLC on May 31, 2023 and is designated Class II. Reason for recall: Screws for osteotomies have contain an undersize screwhead which has the potential to pass through the plate hole. The recall status is ongoing.
Recall details
- Product Type
- Devices
- Report Date
- Jul 19, 2023
- Initiation Date
- May 31, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
Product description
REF: 218-1604, 1.6mm x 4mm Profile Zero Auto-Drive Screw, Rx Only, Sterile
Reason for recall
Screws for osteotomies have contain an undersize screwhead which has the potential to pass through the plate hole.
Distribution pattern
Worldwide - US Nationwide distribution in the states of AZ, CA, FL, GA, KS, KY, MD, MO, MT, NE, NY, TN, TX and the country of SG.
Code information
Lot # 1177027, 1177718, 1177812 / UDI-DI: (01)00845694034258
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2107-2023
- openFDA file date
- Nov 19, 2026
- Loaded by RegDataLab
- Nov 20, 2026
Need the full FDA record for Osteomed, LLC? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.