Z-2108-2020 Class II Terminated

Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

FDA device recall Z-2108-2020 was initiated by Siemens Medical Solutions USA, Inc on Apr 22, 2020 and is designated Class II. Reason for recall: If the C-arm leaves its intended travel path due to a fault within the drivetrain, movements may be impacted or impossible and the area of interest cannot be reached. In the event the C-arms movements… The recall status is terminated (terminated Oct 9, 2020). Affected quantity: 105 systems.

Recall details

Product Type
Devices
Report Date
Jun 3, 2020
Initiation Date
Apr 22, 2020
Termination Date
Oct 9, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
105 systems

Product description

ARTIS Pheno - Interventional Fluoroscopic X-Ray System Model: 10849000 - Product Usage: is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.

Reason for recall

If the C-arm leaves its intended travel path due to a fault within the drivetrain, movements may be impacted or impossible and the area of interest cannot be reached. In the event the C-arms movements are not possible, the system can only be returned to normal operation with the support of a field service engineer

Distribution pattern

US Nationwide distribution.

Code information

Serial Numbers: 164011 164013 164018 164019 164021 164022 164024 164025 164026 164027 164028 164033 164034 164048 164049 164053 164054 164055 164057 164060 164061 164063 164064 164066 164068 164069 164071 164072 164073 164074 164075 164080 164083 164084 164086 164088 164094 164095 164097 164098 164100 164105 164113 164115 164121 164124 164125 164132 164134 164138 164140 164142 164152 164153 164155 164157 164158 164159 164160 164161 164163 164168 164171 164172 164173 164183 164184 164186 164187 164188 164189 164190 164196 164201 164202 164204 164205 164210 164211 164212 164213 164216 164227 164228 164231 164234 164240 164241 164242 164244 164250 164253 164257 164263 164267 164269 164270 164271 164275 164277 164279 164282 164287 164292 164293

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2108-2020
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Siemens Medical Solutions USA, Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Siemens Medical Solutions USA, Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Siemens Medical Solutions USA, Inc →

Source: openFDA record for Z-2108-2020. Report a data issue: hello@regdatalab.com.