Z-2110-2021 Class I Terminated

Recalled by Baxter Healthcare Corporation — Deerfield, IL

Recall Details

Product Type
Devices
Report Date
August 18, 2021
Initiation Date
June 4, 2021
Termination Date
September 25, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
130135 units

Product Description

Spectrum IQ Infusion System with Dose IQ Safety Software, Product Code 3570009. for the controlled administration of fluids.

Reason for Recall

System errors were noted following changes the customers implemented to the configuration of their network and server systems. Gateway Server System performance anomalies resulted in excessive Spectrum IQ pump connectivity errors.

Distribution Pattern

United States (including Puerto Rico), Canada, Bahamas, Bermuda, Guyana, and Trinidad and Tobago.

Code Information

All Serial Numbers