Z-2115-2020 Class II Terminated
FDA device recall Z-2115-2020 was initiated by Inpeco S.A. on Apr 7, 2020 and is designated Class II. Reason for recall: The sensors which activate the safety switches may be affected by delayed activation. In 1% of the cases, the response delay can be up to a maximum delay of 2 seconds, instead of 40ms expected by… The recall status is terminated (terminated Jun 13, 2024). Affected quantity: 2 systems.
Recall details
- Product Type
- Devices
- Report Date
- Jun 3, 2020
- Initiation Date
- Apr 7, 2020
- Termination Date
- Jun 13, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 2 systems
Product description
Reason for recall
Distribution pattern
The products were distributed to the following US States: IL and NY.