Z-2116-2025 Class II Ongoing

Recalled by ICU Medical, Inc. — San Clemente, CA

Recall Details

Product Type
Devices
Report Date
July 23, 2025
Initiation Date
June 3, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1,445,600

Product Description

SwabCap Disinfecting Cap for Needlefree Connectors: SwabSleeve, REF: SCXT3-5-2000; SwabCap, 200-ct. Box, REF: SCXT3-2000

Reason for Recall

Due to a manufacturing issue, disinfecting cap for needle-free connectors may have an incomplete seal between the foil lid and plastic container, which may result in isopropyl alcohol evaporation from the sponge, which may result in inadequate disinfection.

Distribution Pattern

Worldwide - US Nationwide distribution including in the states of TX, SD, ID, LA, IN, CO, OK, MO, MN, OR, AZ, NC, TN, VA, CA, IL, FL, MA, OH, MD, PA, NJ, NY, WA, GA, UT, MI, MS, AR, AL, NV, KS, IA, WV, HI, KY, NM and the country of CA.

Code Information

REF/UDI-DI/Lot(Expiration): SCXT3-5-2000/10887709097107/14258739(1-Dec-26), 14258743(1-Feb-26), 14258744(1-Dec-26), 14264200(1-Dec-26), 14272241(1-Dec-26), 14272248(1-Dec-26), 14287456(1-Dec-26), 14287459(1-Dec-26), 14287480(1-Dec-26), 14287482(1-Dec-26), 14287485(1-Dec-26), 14294832(1-Dec-26), 14272254(1-Dec-26), 14310313(1-Dec-26), 14310425(1-Dec-26), 14323354(1-Dec-26), 14287475(1-Dec-26), 14294840(1-Dec-26), 14321645(1-Jan-27), 14321652(1-Jan-27), 14331169(1-Jan-27), 14332773(1-Jan-27), 14341732(1-Oct-26), 14294825(1-Jan-27), 14303209(1-Jan-27), 14310308(1-Jan-27), 14355404(1-Jan-27), 14355409(1-Jan-27) SCXT3-2000/20887709079162/14258214(1-Jan-27), 14258206(1-Jan-27), 14258208(1-Feb-27).