Z-2124-2026 Class II Ongoing

Recalled by Medline Industries, LP — Northfield, IL

FDA device recall Z-2124-2026 was initiated by Medline Industries, LP on Jan 7, 2026 and is designated Class II. Reason for recall: Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however… The recall status is ongoing. Affected quantity: 240 units.

Recall details

Product Type
Devices
Report Date
May 13, 2026
Initiation Date
Jan 7, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
240 units

Product description

Medline Convenience Kits: 1) DRAPE PACK, Model Number: DYNJ35363A; 2) SICK KIDS-GENERAL CUSTOM SPLIT, Model Number: SPTPCD10005

Reason for recall

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

Distribution pattern

Worldwide - US Nationwide distribution.

Code information

1) DYNJ35363A, UDI-DI: 10884389516461(each), 40884389516462(case), Lot Number: 21BBP020; 2) SPTPCD10005, UDI-DI: 10080196228652(each), 40080196228653(case), Lot Number: 24JBB041

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2124-2026
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2124-2026. Report a data issue: hello@regdatalab.com.