Z-2126-2014 Class II Terminated

Recalled by Baxter Healthcare Corporation — Westlake Village, CA

FDA device recall Z-2126-2014 was initiated by Baxter Healthcare Corporation on Jul 11, 2014 and is designated Class II. Reason for recall: Baxter Healthcare is recalling the FlowCOUPLER Device because there is a possibility that some units may not maintain electrical continuity during handling intra-operatively and post-operatively… The recall status is terminated (terminated May 27, 2015). Affected quantity: 1,315 units.

Recall details

Product Type
Devices
Report Date
Aug 6, 2014
Initiation Date
Jul 11, 2014
Termination Date
May 27, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1,315 units

Product description

FlowCOUPLER. An implantable device that is used to detect blood flow in vessels. Used in the end-to-end anastomosis of veins and arteries normally encountered in microsurgical and vascular reconstructive procedures. The FlowCOUPLER System is intended to detect blood flow and confirm vessel patency intra-operatively and post-operatively at the anastomotic site. Baxter Healthcare.

Reason for recall

Baxter Healthcare is recalling the FlowCOUPLER Device because there is a possibility that some units may not maintain electrical continuity during handling intra-operatively and post-operatively, resulting in a loss of Doppler signal. This may result in the inability to monitor anastomotic patency, with the possibility that blood flow may not be appropriately monitored.

Distribution pattern

Worldwide Distribution - USA (nationwide) and the countries of United Kingdom, Germany, Sweden, Spain, Israel and Denmark.

Code information

Lot Number SPSGT14-02B0027a SPSGT14-04D0005 SPSGT14-04D0023 SPSGT13-10J0016 SPSGT13-11K0041 SPSGT13-12L0008 SPSGT14-01A0009 SPSGT14-02B0026 SPSGT14-04D0015 SPSGT14-04D0026 SPSGT13-10J0017 SPSGT13-11K0039 SPSGT13-11K0042 SPSGT13-12L0015 SPSGT14-01A0028 SPSGT14-02B0028 SPSGT14-02B0033 SPSGT14-04D0014 SPSGT14-04D0027 SPSGT14-05E0005 SPSGT13-10J0018 SPSGT13-11K0040 SPSGT13-12L0007 SPSGT13-12L0012 SPSGT14-01A0027 Part No. 5151-01200-010 5151-01200-010 5151-01200-010 5151-01200-011 5151-01200-011 5151-01200-011 5151-01200-011 5151-01250-010 5151-01250-010 5151-01250-010 5151-01250-011 5151-01250-011 5151-01250-011 5151-01250-011 5151-01250-011 5151-01300-010 5151-01300-010 5151-01300-010 5151-01300-010 5151-01300-010 5151-01300-011 5151-01300-011 5151-01300-011 5151-01300-011 5151-01300-011

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2126-2014
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2126-2014. Report a data issue: hello@regdatalab.com.