Z-2131-2021 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- July 28, 2021
- Initiation Date
- June 29, 2021
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 89,050 total
Product Description
Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 138-28-XX, Acumatch GXL 15 Degree, +5 Lat Liner, 28mm; Novation GXL 10 Deg Liner, G0, 28mm; and Novation GXL Liner, 10 Deg Face, 28mm. Orthopedic hip implant component.
Reason for Recall
Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.
Distribution Pattern
Distribution US nationwide.
Code Information
All serial numbers. UDI-DI numbers: 10885862024589, 10885862024596, 10885862024602, 10885862024619, 10885862024626, 10885862024633, 10885862024664, 10885862024671