Z-2132-2012 Class II Terminated

Recalled by Baxter Healthcare Renal Div — Waukegan, IL

Recall Details

Product Type
Devices
Report Date
August 8, 2012
Initiation Date
July 6, 2012
Termination Date
October 15, 2012
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
N/A

Product Description

Extraneal and/or DIANEAL LOW CALCIUM Peritoneal Dialysis Solution in Plastic Container, PL-146 used with Automated Peritoneal Dialysis System/Automated PD Set with Cassette; Manufactured for Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A. The HomeChoice dialysis system is a personal cycler automated peritoneal dialysis system for pediatric and adult renal patients with fill volumes ranging from 60 mL to 3000 mL. DIANEAL Low Calcium peritoneal dialysis solutions are indicated for use in chronic renal failure patients being maintained on peritoneal dialysis. - and- The Personal Cycler Peritoneal Dialysis System [cycler and ancillary sets] is used for automatic control of dialysate solution exchanges in treatment of renal failure patients undergoing peritoneal dialysis. The Personal Cycler System automatically cycles a prescribed volume of dialysis solution into and out of the peritoneal cavity during the dialysis treatment.

Reason for Recall

Baxter discovered that a Baxter technical service representative was providing incorrect instructions for inspection of solution bags for a blocked frangible prior to connection to the Automated PD Set with Cassette.

Distribution Pattern

Nationwide Distribution includingArizona, California, Ohio, Pennsylvania and Virginia.

Code Information

all lots and serial numbers