Z-2132-2013 Class II Terminated

Recalled by DePuy Spine, Inc. — Raynham, MA

Recall Details

Product Type
Devices
Report Date
September 11, 2013
Initiation Date
April 17, 2013
Termination Date
August 13, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3896 units (Overall quantity distributed appears greater than quantities manufactured due to the distribution pathways)

Product Description

CONFIDENCE SPINAL CEMENT SYSTEM 5CC KIT, Product Code: 2839-05-000 The CONFIDENCE Spinal Cement System is intended for percutaneous delivery of CONFIDENCE Spinal Cement, which is indicated for fixation of pathological fractures of the vertebral body during vertebroplasty or kyphoplasty procedures.

Reason for Recall

During injection of cement, the water in the hydraulic pump leaks past the piston within the pump body resulting in the loss of pressure and inability to continue to inject cement.

Distribution Pattern

Worldwide Distribution - USA including AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IN, KY, LA, MA, MD, MI, MN, MO, NC, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WV, WY; Worldwide: Algeria, Australia, Austria, Belgium, Brazil, Canada, Colombia, Cyprus, Czech Republic, France, Germany, Hong Kong, India, Israel, Italy, Latvia, Luxembourg, Malaysia, Mexico, Netherlands, Norway, Poland, Portugal, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, UAE, and UK

Code Information

HPBBZP, HPDBFZ