Z-2132-2020 Class II Terminated
Recalled by Smith & Nephew, Inc.
— Memphis, TN
FDA device recall Z-2132-2020 was initiated by Smith & Nephew, Inc. on Jan 28, 2020 and is designated Class II. Reason for recall: The spike rail was manufactured with an oversized thru-hole which could potentially cause the dowel pin to fall out and the device to disassemble. The recall status is terminated (terminated Dec 11, 2023). Affected quantity: 200 units.
Recall details
- Product Type
- Devices
- Report Date
- Jun 3, 2020
- Initiation Date
- Jan 28, 2020
- Termination Date
- Dec 11, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 200 units
Product description
JOURNEY II UNI RESECTION PREP TRIAL, LEFT MEDIAL/ RIGHT LATERAL - Product Usage: The JOURNEY II Uni Resection Prep Trial is a reusable surgical instrument used during range of motion trialing to assess joint tension and component alignment during surgery. The provisional moveable spikes (spike rail) aids in provisional fixation of the trial. SIZE 1, CATALOG NUMBER: REF 74035821 SIZE 2, CATALOG NUMBER: REF 74035822 SIZE 3, CATALOG NUMBER: REF 74035823 SIZE 4, CATALOG NUMBER: REF 74035824 SIZE 5, CATALOG NUMBER: REF 74035825 SIZE 6, CATALOG NUMBER: REF 74035826 SIZE 7, CATALOG NUMBER: REF 74035827 SIZE 8, CATALOG NUMBER: REF 74035828 SIZE 9, CATALOG NUMBER: REF 74035829 SIZE 10, CATALOG NUMBER: REF 74035830
Reason for recall
The spike rail was manufactured with an oversized thru-hole which could potentially cause the dowel pin to fall out and the device to disassemble.
Distribution pattern
US Nationwide distribution.
Code information
SIZE 1, CATALOG NUMBER: REF 74035821; LOT/BATCH CODES: 19EM19112R, 19FM05587R, 19FM11174R, 19HM18075, 19HM19648, 19HM21015, 19HM21016, 19HM23242 SIZE 2, CATALOG NUMBER: REF 74035822; LOT/BATCH CODES: 19EM21573R, 19EM21573S, 19FM05589R, 19HM21031, 19HM21033, 19HM21878, 19HM21881 SIZE 3, CATALOG NUMBER: REF 74035823; LOT/BATCH CODES: 19EM21578R, 19HM19009, 19JM09476, 19JM18509, 19JM18525, 19JM21072, 19KM14473, 19KM17780 SIZE 4, CATALOG NUMBER: REF 74035824; LOT/BATCH CODES: 19EM14365R, 19EM14365S, 19FM11173, 19HM18078, 19HM19659, 19HM21861, 19JM21075, 19KM15573, 19KM1557 SIZE 5, CATALOG NUMBER: REF 74035825; LOT/BATCH CODES: 19EM14367R, 19HM18079, 19HM21855, 19KM12683 SIZE 6, CATALOG NUMBER: REF 74035826; LOT/BATCH CODES: 19EM24480R, 19HM18080, 19HM21856, 19KM21512 SIZE 7, CATALOG NUMBER: REF 74035827; LOT/BATCH CODES: 19EM14371R, 19FM01063R, 19FM11177R, 19HM18335, 19HM21873, 19HM21874, 19KM17781 SIZE 8, CATALOG NUMBER: REF 74035828; LOT/BATCH CODES: 19EM24498R, 19EM24498S, 19FM05591R, 19FM11178R, 19FM11178S, 19JM07464, 19JM07465, 19KM07278, 19KM12699 SIZE 9, CATALOG NUMBER: REF 74035829; LOT/BATCH CODES: 19EM21582R, 19FM11179R, 19GM21787, 19JM07467, 19KM07276, 19KM15575, 19KM15576 SIZE 10, CATALOG NUMBER: REF 74035830; LOT/BATCH CODES: 19EM14372R, 19FM11180, 19HM18254, 19JM07468, 19JM07470, 19JM07472, 19KM05610, 19KM15571
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2132-2020
- openFDA file date
- Nov 25, 2026
- Loaded by RegDataLab
- Nov 26, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Smith & Nephew, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Smith & Nephew, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about Smith & Nephew, Inc. →
Use it in ChatGPT
Get a free API key
How the connector works