Z-2133-2014 Class III Terminated

Recalled by Siemens Healthcare Diagnostics, Inc. — Newark, DE

FDA device recall Z-2133-2014 was initiated by Siemens Healthcare Diagnostics, Inc. on May 27, 2014 and is designated Class III. Reason for recall: There is a misassembly issue with the photometer installed on Dimension¿ Integrated Chemistry Systems instruments. Through internal investigation by Siemens Healthcare Diagnostics has confirmed that… The recall status is terminated (terminated Dec 15, 2014). Affected quantity: 15.

Recall details

Product Type
Devices
Report Date
Aug 6, 2014
Initiation Date
May 27, 2014
Termination Date
Dec 15, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
15

Product description

Dimension¿ Integrated Chemistry Systems intended to measure a variety of analytes in human body fluids.

Reason for recall

There is a misassembly issue with the photometer installed on Dimension¿ Integrated Chemistry Systems instruments. Through internal investigation by Siemens Healthcare Diagnostics has confirmed that a limited number of Dimension photometers were misassembled. These photometers are performing within specifications, but may lead to a shortened life of the source lamp.

Distribution pattern

FL LA AR IA OH OH WI IL LA KS MT

Code information

DE271317, DE271320, DE271323, DE271324, DE271326, DE271327, DE271328, DE271332 DE271341, DE271343, DE271325, 12252589, 2004082898, 2004082899, 224719-AX

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2133-2014
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2133-2014. Report a data issue: hello@regdatalab.com.