Z-2133-2014 Class III Terminated
FDA device recall Z-2133-2014 was initiated by Siemens Healthcare Diagnostics, Inc. on May 27, 2014 and is designated Class III. Reason for recall: There is a misassembly issue with the photometer installed on Dimension¿ Integrated Chemistry Systems instruments. Through internal investigation by Siemens Healthcare Diagnostics has confirmed that… The recall status is terminated (terminated Dec 15, 2014). Affected quantity: 15.
Recall details
- Product Type
- Devices
- Report Date
- Aug 6, 2014
- Initiation Date
- May 27, 2014
- Termination Date
- Dec 15, 2014
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 15
Product description
Dimension¿ Integrated Chemistry Systems intended to measure a variety of analytes in human body fluids.
Reason for recall
There is a misassembly issue with the photometer installed on Dimension¿ Integrated Chemistry Systems instruments. Through internal investigation by Siemens Healthcare Diagnostics has confirmed that a limited number of Dimension photometers were misassembled. These photometers are performing within specifications, but may lead to a shortened life of the source lamp.
Distribution pattern
FL LA AR IA OH OH WI IL LA KS MT
Code information
DE271317, DE271320, DE271323, DE271324, DE271326, DE271327, DE271328, DE271332 DE271341, DE271343, DE271325, 12252589, 2004082898, 2004082899, 224719-AX