Z-2135-2012 Class II Terminated

Recalled by Stryker Instruments Div. Of Stryker Corporation — Portage, MI

FDA device recall Z-2135-2012 was initiated by Stryker Instruments Div. Of Stryker Corporation on Jul 17, 2012 and is designated Class II. Reason for recall: Stryker has received 6 complaints that the 5mm needle was placed in 10 mm package and 10 mm needle was placed in the 5 mm package. If the clinician does not recognize that the active tip on the… The recall status is terminated (terminated May 2, 2013). Affected quantity: 226.

Recall details

Product Type
Devices
Report Date
Aug 8, 2012
Initiation Date
Jul 17, 2012
Termination Date
May 2, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
226

Product description

RF Cannula, 100 mm,Curved 10 MM, 20 G, REF 0406-630-125 RX only, Disposable ( 10/pk) Product Usage: The Stryker RF Cannulae, in combination with the Stryker RF Generator/Multi- Gen and electrodes, are intended for coagulation of soft tissue in orthopedic, arthroscopic, spinal, and neurosurgical applications.

Reason for recall

Stryker has received 6 complaints that the 5mm needle was placed in 10 mm package and 10 mm needle was placed in the 5 mm package. If the clinician does not recognize that the active tip on the cannula is 10 mm instead of 5mm and the cannula is used in a cervical procedure there is a potential for a thermal skin lesion due to a larger lesion creation. If the 10 mm active tip cannula, as opposed

Distribution pattern

Worldwide Distribution - US Nationwide and the countries of England, France, Netherlands, and Switzerland.

Code information

REF 0406-630-125, lot 1000037535

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2135-2012
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2135-2012. Report a data issue: hello@regdatalab.com.