Z-2140-2018 Class III Terminated

Recalled by Alere San Diego Inc. DBA Immunalysis Corporation — Pomona, CA

Recall Details

Product Type
Devices
Report Date
June 20, 2018
Initiation Date
November 28, 2017
Termination Date
January 29, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
37 kits

Product Description

lmmunalysis EDDP Urine Control Set 1 The EDDP Urine Controls (High 125 ng/ml and Low 75 ng/ml) are used as control materials in the ammonolysis EDDP Specific Urine Enzyme Immunoassay, which has a cutoff of 100 ng/ml. This test provides only a preliminary analytical test result and a more specific alternate chemical method, such as Gas Chromatography/Mass Spectrometry or Liquid Chromatography/Tandem Mass Spectrometry, must be used to obtain a confirmed analytical result.

Reason for Recall

The High (125 ng/ml) Control does not meet specification - it has been confirmed to produce values below 100 ng/mL when run on validated clinical chemistry analyzers, which is below the 100 ng/mL cutoff of the EDDP Specific Urine Enzyme Immunoassay with which this control is utilized.

Distribution Pattern

US Distribution to the states of : CA, CO, CT, FL, LA, MD, NC, NJ, OH, OK, TN, TX, WI, and WV

Code Information

Part Number: 3006-5, Lot Number: EK16135