Z-2140-2020 Class II Terminated
Recalled by Inpeco S.A.
— Lugano, N/A
FDA device recall Z-2140-2020 was initiated by Inpeco S.A. on Apr 7, 2020 and is designated Class II. Reason for recall: The identified problem is an erroneous association between the carrier and the sample tube caused by a communication error between the firmware of the Interface Modules and the Automation software… The recall status is terminated (terminated Feb 6, 2024). Affected quantity: 37 systems.
Recall details
- Product Type
- Devices
- Report Date
- Jun 3, 2020
- Initiation Date
- Apr 7, 2020
- Termination Date
- Feb 6, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 37 systems
Product description
FlexLab Automation System with Advia2120LAS Interface Module (P/N FLX-219-00, FLX-219-10), ImmunoCAP 1000 Interface Module (P/N FLX-226-01, FLX-226-10), or StaRRsed Interface Module (P/N FLX-268-00)is a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The systems consolidate multiple Analytical instruments into a unified workstation.
Reason for recall
The identified problem is an erroneous association between the carrier and the sample tube caused by a communication error between the firmware of the Interface Modules and the Automation software. This problem can occur only when one of these Interface Modules is put off-line after a carrier in their secondary lane is physically returned on the main track and then put back on-line when the carrier is used to transport another tube.
Distribution pattern
US Nationwide including in the states of IL, NY.
Code information
Serial Numbers: FLX.005 FLX.006 FLX.011 FLX.0117 FLX.0121 FLX.0124 FLX.0126 FLX.0129 FLX.0133 FLX.0136 FLX.0139 FLX.0140 FLX.0147 FLX.0149 FLX.0153 FLX.0157 FLX.016 FLX.0165 FLX.0168 FLX.0173 FLX.0189 FLX.0191 FLX.0192 FLX.022 FLX.030 FLX.032 FLX.035 FLX.036 FLX.037 FLX.050 FLX.056 FLX.060 FLX.062 FLX.065 FLX.082 FLX.096 SHD.002
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2140-2020
- openFDA file date
- Nov 24, 2026
- Loaded by RegDataLab
- Nov 25, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Inpeco S.A.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.
Get a free API key
How the connector works