Z-2141-2020 Class II Terminated
FDA device recall Z-2141-2020 was initiated by Inpeco S.A. on Apr 7, 2020 and is designated Class II. Reason for recall: The identified problem is an erroneous association between the carrier and the sample tube caused by a communication error between the firmware of the Interface Modules and the Automation software… The recall status is terminated (terminated Feb 6, 2024). Affected quantity: 3 systems.
Recall details
- Product Type
- Devices
- Report Date
- Jun 3, 2020
- Initiation Date
- Apr 7, 2020
- Termination Date
- Feb 6, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 3 systems
Product description
Reason for recall
Distribution pattern
US Nationwide including in the states of IL, NY.