Z-2150-2024 Class II Ongoing

Recalled by Outset Medical, Inc. — San Jose, CA

Recall Details

Product Type
Devices
Report Date
June 26, 2024
Initiation Date
April 30, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
431

Product Description

TabloCart with Prefiltration Drawer PN-0006813, with power cords: PN-0005746 (10-foot), PN-0006578 (15-foot), PN-0005747 (20-foot)10-, 15- and 20-foot cords respectively), used with the Tablo Hemodialysis System.

Reason for Recall

A cart with prefiltration drawer, an optional hemodialysis system accessory, has a cord with a female connector that has internal contact pins that may become damaged that could create localized heat, sparks, smoke, melting or charring (thermal event) of the female connector or the associated connector power entry module, which may lead to transient delay in treatment or blistering to the user.

Distribution Pattern

US Nationwide distribution in the states of FL, CA, NC, GA, MD, DE, AK, OK, MA, WV, IN, KY, AZ, LA, NE, KS, NJ, CT, NM, CO, SC, TX, PA, NY, OH.

Code Information

UDI-DI: 00850001011303. All PN-0006813 serial numbers with power cords: PN-0005746, PN-0006578, PN-0005747