Z-2151-2023 Class I Ongoing

Recalled by Datascope Corp. — Mahwah, NJ

FDA device recall Z-2151-2023 was initiated by Datascope Corp. on Jun 5, 2023 and is designated Class I. Reason for recall: An unexpected shutdown of the IABP may occur due to failures of tantalum capacitors in the Power Management Board and/or Solenoid Board. This issue may lead to an unexpected interruption of therapy. The recall status is ongoing. Affected quantity: 9174 (4586 US; 4588 OUS) in total.

Recall details

Product Type
Devices
Report Date
Jul 26, 2023
Initiation Date
Jun 5, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
9174 (4586 US; 4588 OUS) in total

Product description

Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model numbers 0998-00-0800-31 0998-00-0800-32 0998-00-0800-33 0998-00-0800-34 0998-00-0800-35 0998-00-0800-45 0998-00-0800-52 0998-00-0800-53 0998-00-0800-55 0998-00-0800-65 0998-UC-0800-31* 0998-UC-0800-33* 0998-UC-0800-52* 0998-UC-0800-53* 0998-00-0800-55* *Not US distribution

Reason for recall

An unexpected shutdown of the IABP may occur due to failures of tantalum capacitors in the Power Management Board and/or Solenoid Board. This issue may lead to an unexpected interruption of therapy.

Distribution pattern

Worldwide Distribution- US (Nationwide) and OUS (foreign) countries of: ALBANIA, ALGERIA, ANGOLA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, AZERBAIJAN , BAHRAIN, BANGLADESH, BARBADOS, BELARUS, BELGIUM, BOSNIA and HERZEGOVINA, BRAZIL, BRUNEI, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, C¿te d'Ivoire, CROATIA, CZECH REPUBLIC, ECUADOR, EGYPT, EL SALVADOR, ESTONIA, FINLAND, FRANCE, GERMANY, GHANA, GUATEMALA, HONG KONG, HUNGARY, ICELAND, INDIA, INDONESIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KENYA, KOSOVO, KUWAIT, LATVIA, LEBANON, LIBYA, LITHUANIA, LUXEMBOURG, MACEDONIA, MALAYSIA, MEXICO, MOROCCO, MYANMAR, NETHERLANDS, NEW ZEALAND, NIGERIA, NORWAY, OMAN, PAKISTAN, PALESTINE, PERU, PHILIPPINES, POLAND, PORTUGAL, QATAR, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, SYRIA, TAIWAN, TANZANIA, THAILAND, TRINIDAD AND TOBAGO, TUNISIA, TURKEY, TURKMENISTAN, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM, UZBEKISTAN, VIETNAM, YEMEN, ZAMBIA

Code information

All Unit Serial Numbers Model UDI 0998-00-0800-31 10607567109053 0998-00-0800-32 10607567111117 0998-00-0800-33 10607567109008 0998-00-0800-34 10607567111940 0998-00-0800-35 10607567109107 0998-00-0800-45 10607567108421 0998-00-0800-52 10607567108438 0998-00-0800-53 10607567108391 0998-00-0800-55 10607567108414 0998-00-0800-65 10607567113432

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2151-2023
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2151-2023. Report a data issue: hello@regdatalab.com.