Z-2160-2013 Class II Terminated

Recalled by Villa Radiology Systems LLC — Oxford, CT

Recall Details

Product Type
Devices
Report Date
September 25, 2013
Initiation Date
August 2, 2013
Termination Date
July 25, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
22 units within US

Product Description

Juno DRF System. Mobile X-ray system These are general, multi-function x-ray systems capable of routine radiographs and general fluoroscopy procedures

Reason for Recall

It has been discovered that the Air Kerma rate information is inconsistent in Continuous Fluoro modes 30x30 (12"x12") and 15x15 (6"x6). This could result in a deviation of the displayed values of Air Kerma Rate and Cumulated Air Kerma from the actual values greater then +/-35% and then in a failure to comply with the requirements of 21 CFR 1020.32(k)(6

Distribution Pattern

US Distribution including the states of AZ, MI, IN, UT, CT, AL, NY, PA, IL, MD, CA, MA, IA, OH, MS and MO.

Code Information

PN 709020