Z-2161-2016 Class II Terminated

Recalled by Stryker Neurovascular — Fremont, CA

FDA device recall Z-2161-2016 was initiated by Stryker Neurovascular on Jun 6, 2016 and is designated Class II. Reason for recall: Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been… The recall status is terminated (terminated Mar 3, 2017). Affected quantity: 45,302 (8,188 in the U.S.) - total, all model numbers.

Recall details

Product Type
Devices
Report Date
Jul 20, 2016
Initiation Date
Jun 6, 2016
Termination Date
Mar 3, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
45,302 (8,188 in the U.S.) - total, all model numbers

Product description

Target Detachable Coils: TARGET 360 NANO 1 MM X 3 CM MODEL Number: M0035421030 Neurology: Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.

Reason for recall

Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased occurrence of coil stretching.

Distribution pattern

Worldwide Distribution - US (nationwide) and Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italia, Japan,Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Mexico,Netherlands, Norway, Peru, Philippines, Poland, Portugal, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom, United States, Uruguay, and Vietnam.

Code information

MODEL Number: M0035421030: Lot Number/Expiration Date: 18604566 30-Sep-18 18488920 31-Jul-18 18807744 31-Jan-19 18807570 31-Jan-19 18677403 31-Oct-18 18666895 31-Oct-18 18654009 31-Oct-18 18647637 31-Oct-18 18629287 31-Oct-18 18542421 31-Aug-18 18554251 31-Aug-18 18469669 30-Jun-18 18462274 30-Jun-18 18462217 30-Jun-18 18393386 30-Apr-18 18814099 31-Jan-19 18788265 31-Jan-19 18769352 31-Dec-18 18714536 30-Nov-18 18701840 30-Nov-18 18677577 30-Nov-18 18677461 30-Nov-18 18629171 31-Oct-18 18604508 30-Sep-18 18582681 30-Sep-18 18535009 31-Aug-18 18426357 30-Jun-18 18714768 30-Nov-18 18714652 30-Nov-18 18629229 31-Oct-18 18554193 31-Aug-18 18701898 30-Nov-18 18730274 30-Nov-18 18677519 30-Nov-18 18653835 31-Oct-18 18831481 31-Jan-19 18769468 31-Dec-18 18714304 30-Nov-18 18807918 31-Jan-19 18714420 30-Nov-18 18654067 31-Oct-18 18638376 31-Oct-18 18590876 30-Sep-18 18604450 30-Sep-18 18488978 31-Jul-18 18469612 30-Jun-18 18831539 31-Jan-19 18788439 31-Jan-19 18736862 31-Dec-18 18736746 30-Nov-18 18666837 31-Oct-18 18666779 31-Oct-18 18653893 31-Oct-18 18604392 30-Sep-18 18542479 31-Aug-18 18931539 31-Mar-19 18901428 28-Feb-19 19048197 31-May-19 18945290 31-Mar-19 18931771 31-Mar-19 18846927 31-Jan-19 18909826 31-Mar-19 18887689 28-Feb-19 18887573 28-Feb-19 18847159 28-Feb-19 18976078 30-Apr-19 19025545 30-Apr-19 19043431 31-May-19 18975900 30-Apr-19 18975989 30-Apr-19 19067208 31-May-19 18954139 31-Mar-19 18961180 31-Mar-19 18961060 31-Mar-19 18945174 31-Mar-19 18931307 31-Mar-19 18847043 31-Jan-19 19025429 30-Apr-19 19048139 31-May-19 18945406 31-Mar-19 18901486 28-Feb-19 18909768 28-Feb-19 19043373 31-May-19 19012218 30-Apr-19 19012392 30-Apr-19 18976167 30-Apr-19 18854389 28-Feb-19 19012044 30-Apr-19.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2161-2016
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Stryker Neurovascular? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Stryker Neurovascular: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Stryker Neurovascular →

Source: openFDA record for Z-2161-2016. Report a data issue: hello@regdatalab.com.