Z-2171-2016 Class II Terminated

Recalled by Stryker Neurovascular — Fremont, CA

FDA device recall Z-2171-2016 was initiated by Stryker Neurovascular on Jun 6, 2016 and is designated Class II. Reason for recall: Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been… The recall status is terminated (terminated Mar 3, 2017). Affected quantity: 45,302 (8,188 in the U.S.) - total, all model numbers.

Recall details

Product Type
Devices
Report Date
Jul 20, 2016
Initiation Date
Jun 6, 2016
Termination Date
Mar 3, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
45,302 (8,188 in the U.S.) - total, all model numbers

Product description

Target Detachable Coils: TARGET HELICAL NANO 1.5 MM X 4 CM MODEL Number:M0035431540 Neurology: Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.

Reason for recall

Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased occurrence of coil stretching.

Distribution pattern

Worldwide Distribution - US (nationwide) and Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italia, Japan,Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Mexico,Netherlands, Norway, Peru, Philippines, Poland, Portugal, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom, United States, Uruguay, and Vietnam.

Code information

MODEL Number: M0035431540: Lot Number/Expiration Date: 18843121 28-Feb-19 18842254 28-Feb-19 18752256 31-Dec-18 18637622 31-Oct-18 18531974 31-Aug-18 18450202 30-Jun-18 18765704 31-Dec-18 18749439 31-Dec-18 18748627 31-Dec-18 18546175 31-Aug-18 18546117 31-Aug-18 18556378 31-Aug-18 18538344 31-Aug-18 18749033 31-Dec-18 18602307 30-Sep-18 18532032 31-Jul-18 18802864 31-Jan-19 18797083 31-Jan-19 18794216 31-Jan-19 18812672 31-Jan-19 18637738 31-Oct-18 18607230 31-Oct-18 18637680 31-Oct-18 18449404 30-Jun-18 18794100 31-Jan-19 18776421 31-Dec-18 18602365 30-Sep-18 18581643 30-Sep-18 18801549 31-Jan-19 18737036 31-Dec-18 18602249 30-Sep-18 18486380 31-Jul-18 18486322 31-Jul-18 19086644 31-May-19 18941592 31-Mar-19 19086586 31-May-19 18973495 30-Apr-19 19004069 30-Apr-19 18859100 28-Feb-19 18972803 30-Apr-19 19056273 31-May-19 19062764 31-May-19 19003895 30-Apr-19 18973149 30-Apr-19 18956009 31-Mar-19 18893301 28-Feb-19 18894059 28-Feb-19 18860564 28-Feb-19 19003547 30-Apr-19 18955103 31-Mar-19 18942056 31-Mar-19 18886614 28-Feb-19 18894438 28-Feb-19 18941128 31-Mar-19 19083403 31-May-19 18862068 28-Feb-19 18885200 28-Feb-19 19003721 30-Apr-19 19064636 31-May-19.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2171-2016
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Stryker Neurovascular? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Stryker Neurovascular: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Stryker Neurovascular →

Source: openFDA record for Z-2171-2016. Report a data issue: hello@regdatalab.com.