Z-2175-2012 Class III Terminated
FDA device recall Z-2175-2012 was initiated by Siemens Healthcare Diagnostics, Inc. on Jul 17, 2012 and is designated Class III. Reason for recall: Siemens has received complaints of customers experiencing slopes >1.05 when calibrating Flex(R) reagent cartridge lot 12095AD. This results in a slope failure and prevents the VANC assay from being… The recall status is terminated (terminated Jan 14, 2014). Affected quantity: 1,894 cartons.
Recall details
- Product Type
- Devices
- Report Date
- Aug 15, 2012
- Initiation Date
- Jul 17, 2012
- Termination Date
- Jan 14, 2014
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1,894 cartons
Product description
Reason for recall
Distribution pattern
Worldwide Distribution -- US, including Washington, DC and the states of AL, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, and WV, and the countries of Canada, New Zealand, Germany, and Puerto Rico.
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2175-2012