Z-2175-2016 Class II Terminated

Recalled by Stryker Neurovascular — Fremont, CA

FDA device recall Z-2175-2016 was initiated by Stryker Neurovascular on Jun 6, 2016 and is designated Class II. Reason for recall: Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been… The recall status is terminated (terminated Mar 3, 2017). Affected quantity: 45,302 (8,188 in the U.S.) - total, all model numbers.

Recall details

Product Type
Devices
Report Date
Jul 20, 2016
Initiation Date
Jun 6, 2016
Termination Date
Mar 3, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
45,302 (8,188 in the U.S.) - total, all model numbers

Product description

Target Detachable Coils: TARGET 360 NANO 3MM X 4CM MODEL Number:M0035443040 Neurology: Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.

Reason for recall

Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased occurrence of coil stretching.

Distribution pattern

Worldwide Distribution - US (nationwide) and Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italia, Japan,Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Mexico,Netherlands, Norway, Peru, Philippines, Poland, Portugal, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom, United States, Uruguay, and Vietnam.

Code information

MODEL Number:M0035443040 : Lot Number/Expiration Date: 18442838 30-Jun-18 18442388 30-Jun-18 18490600 31-Jul-18 18831276 31-Jan-19 18760116 31-Dec-18 18760119 31-Dec-18 18584035 30-Sep-18 18470827 31-Jul-18 18442358 30-Jun-18 18830247 31-Jan-19 18470767 30-Jun-18 18470857 31-Jul-18 18800019 31-Jan-19 18760117 31-Dec-18 18568317 31-Aug-18 18462748 30-Jun-18 18443108 30-Jun-18 18442178 30-Jun-18 18568224 31-Aug-18 18490538 31-Jul-18 18462838 30-Jun-18 18568193 31-Aug-18 18584004 30-Sep-18 18490662 31-Jul-18 18442688 30-Jun-18 19066241 31-May-19 19035290 31-May-19 18902944 28-Feb-19 18877825 28-Feb-19 18973721 30-Apr-19 18846664 31-Jan-19 19081717 31-May-19 19026562 30-Apr-19 18973029 30-Apr-19 18876957 28-Feb-19 18905865 28-Feb-19 19027086 30-Apr-19 19080897 31-May-19 19063669 31-May-19 18845906 31-Jan-19 18845527 31-Jan-19 19026038 30-Apr-19 18902302 28-Feb-19 18846285 31-Jan-19 19081102 31-May-19 18901660 28-Feb-19.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2175-2016
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Stryker Neurovascular? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Stryker Neurovascular: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Stryker Neurovascular →

Source: openFDA record for Z-2175-2016. Report a data issue: hello@regdatalab.com.