Z-2182-2023 Class II Ongoing

Recalled by MEDLINE INDUSTRIES, LP - Northfield — Northfield, IL

Recall Details

Product Type
Devices
Report Date
July 26, 2023
Initiation Date
May 18, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1380 units

Product Description

Medline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) UNIVERSAL INSERTION TRAY, Model Number CVI4325; b) PPE BUNDLE KIT, Model Number DT13351A

Reason for Recall

Probe covers for use during diagnostic ultrasound procedures may have an inadequate barrier at the seams.

Distribution Pattern

Worldwide Distribution: US (Nationwide) and OUS (foreign) United Arab Emirates, Singapore, Lebanon, China Qatar, Costa Rica, India, Canada, Panama & Saudi Arabia.

Code Information

all lots within expiry, distributed from Medline between December 2017 - May 2023: a) Model Number CVI4325, UDI/DI (EA) 10653160307037, UDI/DI (CS) 40653160307038; b) Model Number DT13351A, UDI/DI (EA) 10653160325581, UDI/DI (CS) 40653160325582