Z-2185-2023 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- July 26, 2023
- Initiation Date
- May 18, 2023
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 3975 units
Product Description
Medline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: UNIVERSAL DRESSING & PPE KIT, Model Number DT22065
Reason for Recall
Probe covers for use during diagnostic ultrasound procedures may have an inadequate barrier at the seams.
Distribution Pattern
Worldwide Distribution: US (Nationwide) and OUS (foreign) United Arab Emirates, Singapore, Lebanon, China Qatar, Costa Rica, India, Canada, Panama & Saudi Arabia.
Code Information
all lots within expiry, distributed from Medline between December 2017 - May 2023: Model Number DT22065, UDI/DI (EA) 10653160326601, UDI/DI (CS) 40653160326602