Z-2194-2013 Class II Terminated

Recalled by Ormco/Sybronendo — Glendora, CA

FDA device recall Z-2194-2013 was initiated by Ormco/Sybronendo on Jul 12, 2013 and is designated Class II. Reason for recall: Ormco Corporation is voluntarily recalling one lot of 8 MM Vector TAS Screws because some screws of this lot were machined undersized at the top of the threads may potentially cause a weakening of the… The recall status is terminated (terminated Sep 11, 2014). Affected quantity: 4,786.

Recall details

Product Type
Devices
Report Date
Sep 18, 2013
Initiation Date
Jul 12, 2013
Termination Date
Sep 11, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4,786

Product description

Brand Name: Vector TAS 8mm Screw. The Device Listing Number is E670535. Vector T AS 8mm Screw is intended to provide a fixed anchorage point for attachment of orthodontic appliances to facilitate the orthodontic movement of teeth. It is used temporarily and removed after orthodontic treatment has been completed. It is surgically placed in the bone of the upper or lower jaw arches to provide support for orthodontic devices.

Reason for recall

Ormco Corporation is voluntarily recalling one lot of 8 MM Vector TAS Screws because some screws of this lot were machined undersized at the top of the threads may potentially cause a weakening of the screw. Use of the undersized product may result in breakage of the screw during insertion or removal when torqued, possibly requiring surgical removal of any remaining pieces.

Distribution pattern

Worldwide Distribution - USA (nationwide) including the countries of Canada, Australia, New Zealand, EU and ROW consignees.

Code information

Lot Number 112911: 8 MM Vector TAS Screw, Part Number 601-0021 Lot Number 112911: Vector TAS Starter Kit (10 Patient), Part Number 601-0010

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2194-2013
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2194-2013. Report a data issue: hello@regdatalab.com.