Z-2206-2015 Class II Terminated

Recalled by American Science & Engineering, Inc. — Billerica, MA

Recall Details

Product Type
Devices
Report Date
August 26, 2015
Initiation Date
May 22, 2015
Termination Date
February 22, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
125 systems

Product Description

American Science & Engineering Inc,Gemini Cabinet X-ray system.

Reason for Recall

During factory testing, we discovered that Gemini systems could, in rare circumstances, experience a fault that leads to a state in which X-rays are being produced but the X-ray indicator lights (located on the operator's console and on both ends of the tunnel) are not illuminated.

Distribution Pattern

Worldwide Distribution - US Distribution to the states of : NY, FL, TN, PA, NJ, DE, ME, TX, CT, MS, NC, MD, MO, VA, DC, NV and GA., and to the countries of : Saudi Arabia, Indonesia, Canada, Netherlands, Hong Kong, Lebanon and Kazakhstan.

Code Information

Gemini systems