Z-2206-2015 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- August 26, 2015
- Initiation Date
- May 22, 2015
- Termination Date
- February 22, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 125 systems
Product Description
American Science & Engineering Inc,Gemini Cabinet X-ray system.
Reason for Recall
During factory testing, we discovered that Gemini systems could, in rare circumstances, experience a fault that leads to a state in which X-rays are being produced but the X-ray indicator lights (located on the operator's console and on both ends of the tunnel) are not illuminated.
Distribution Pattern
Worldwide Distribution - US Distribution to the states of : NY, FL, TN, PA, NJ, DE, ME, TX, CT, MS, NC, MD, MO, VA, DC, NV and GA., and to the countries of : Saudi Arabia, Indonesia, Canada, Netherlands, Hong Kong, Lebanon and Kazakhstan.
Code Information
Gemini systems