Z-2212-2013 Class II Terminated
Recalled by Zeppessis Reprocessing, LLC
— Garden City, ID
FDA device recall Z-2212-2013 was initiated by Zeppessis Reprocessing, LLC on Aug 16, 2013 and is designated Class II. Reason for recall: Zeppessis' reprocessed single-use manifolds (filters) for the Stryker Neptune 1 and Neptune 2 Waste Management Systems are recalled due to lacking 510K and and without a validated cleaning protocol… The recall status is terminated (terminated Nov 8, 2013). Affected quantity: undetermined the distributed amounts.
Recall details
- Product Type
- Devices
- Report Date
- Sep 25, 2013
- Initiation Date
- Aug 16, 2013
- Termination Date
- Nov 8, 2013
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- undetermined the distributed amounts
Product description
Zeppessis' reprocessed single-use manifolds (filters) for the Stryker Neptune 1 and Neptune 2 Waste Management Systems: The Neptune Waste management system is intended to be used in the Operating Room, Surgical Centers and Doctor's Offices to collect and dispose of surgical fluid waste as well as collect smoke generated from electrocautory or laser devices. ***The single-port manifold for the Stryker Neptune 1 (part # 700-015-000) can be used with the: Neptune 1 Gold Rover or the Neptune Bronze Rover as an accessory to Class 2 Air-handling apparatus for a surgical operating room (Sec. 878.5070); or with the Neptune 1 Silver Rover as an accessory to Class 2 Powered suction pump. ***four-port manifold for the Stryker Neptune 1 (part # 700-020-000) can be used with the: Neptune 1 Gold Rover or the Neptune Bronze Rover as an accessory to Class 2 Air-handling apparatus for a surgical operating room (Sec. 878.5070); or with the Neptune 1 Silver Rover as an accessory to Class 2 Powered suction pump. ***single-port manifold for the Stryker Neptune 2 (part # 702-025-000) can be used with the Neptune 2 Rover Ultra as an accessory to Class 2 Powered suction pump. ***four-port manifold for the Stryker Neptune 2 (part # 702-020-000) can be used with the Neptune 2 Rover Ultra as an accessory to Class 2 Powered suction pump.
Reason for recall
Zeppessis' reprocessed single-use manifolds (filters) for the Stryker Neptune 1 and Neptune 2 Waste Management Systems are recalled due to lacking 510K and and without a validated cleaning protocol for reprocessing the used manifolds.
Distribution pattern
Distributed in the states of California, Idaho, Oregon, and Washington.
Code information
-Single-port manifold for the Stryker Neptune 1 (part # 700-015-000); -Four-port manifold for the Stryker Neptune 1 (part # 700-020-000); -Single-port manifold for the Stryker Neptune 2 (part # 702-025-000); -Four-port manifold for the Stryker Neptune 2 (part # 702-020-000).
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2212-2013
- openFDA file date
- Nov 24, 2026
- Loaded by RegDataLab
- Nov 25, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.