Z-2212-2013 Class II Terminated

Recalled by Zeppessis Reprocessing, LLC — Garden City, ID

FDA device recall Z-2212-2013 was initiated by Zeppessis Reprocessing, LLC on Aug 16, 2013 and is designated Class II. Reason for recall: Zeppessis' reprocessed single-use manifolds (filters) for the Stryker Neptune 1 and Neptune 2 Waste Management Systems are recalled due to lacking 510K and and without a validated cleaning protocol… The recall status is terminated (terminated Nov 8, 2013). Affected quantity: undetermined the distributed amounts.

Recall details

Product Type
Devices
Report Date
Sep 25, 2013
Initiation Date
Aug 16, 2013
Termination Date
Nov 8, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
undetermined the distributed amounts

Product description

Zeppessis' reprocessed single-use manifolds (filters) for the Stryker Neptune 1 and Neptune 2 Waste Management Systems: The Neptune Waste management system is intended to be used in the Operating Room, Surgical Centers and Doctor's Offices to collect and dispose of surgical fluid waste as well as collect smoke generated from electrocautory or laser devices. ***The single-port manifold for the Stryker Neptune 1 (part # 700-015-000) can be used with the: Neptune 1 Gold Rover or the Neptune Bronze Rover as an accessory to Class 2 Air-handling apparatus for a surgical operating room (Sec. 878.5070); or with the Neptune 1 Silver Rover as an accessory to Class 2 Powered suction pump. ***four-port manifold for the Stryker Neptune 1 (part # 700-020-000) can be used with the: Neptune 1 Gold Rover or the Neptune Bronze Rover as an accessory to Class 2 Air-handling apparatus for a surgical operating room (Sec. 878.5070); or with the Neptune 1 Silver Rover as an accessory to Class 2 Powered suction pump. ***single-port manifold for the Stryker Neptune 2 (part # 702-025-000) can be used with the Neptune 2 Rover Ultra as an accessory to Class 2 Powered suction pump. ***four-port manifold for the Stryker Neptune 2 (part # 702-020-000) can be used with the Neptune 2 Rover Ultra as an accessory to Class 2 Powered suction pump.

Reason for recall

Zeppessis' reprocessed single-use manifolds (filters) for the Stryker Neptune 1 and Neptune 2 Waste Management Systems are recalled due to lacking 510K and and without a validated cleaning protocol for reprocessing the used manifolds.

Distribution pattern

Distributed in the states of California, Idaho, Oregon, and Washington.

Code information

-Single-port manifold for the Stryker Neptune 1 (part # 700-015-000); -Four-port manifold for the Stryker Neptune 1 (part # 700-020-000); -Single-port manifold for the Stryker Neptune 2 (part # 702-025-000); -Four-port manifold for the Stryker Neptune 2 (part # 702-020-000).

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2212-2013
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Zeppessis Reprocessing, LLC? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2212-2013. Report a data issue: hello@regdatalab.com.