Z-2220-2023 Class II Ongoing

Recalled by Datascope Corp. — Fairfield, NJ

FDA device recall Z-2220-2023 was initiated by Datascope Corp. on Jun 23, 2023 and is designated Class II. Reason for recall: During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral… The recall status is ongoing. Affected quantity: 321,609 total kits.

Recall details

Product Type
Devices
Report Date
Aug 2, 2023
Initiation Date
Jun 23, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
321,609 total kits

Product description

MEGA 8Fr. IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number MEGA 8Fr. 50cc IAB WITH STATLOCK D684-00-0296-01 MEGA 8Fr. 50cc IAB WITH STATLOCK & APA D684-00-0296-02 MEGA 8Fr. 50cc IAB WITH ACCESSORIES (W/O STATLOCK) D684-00-0296-03 MEGA 8Fr. 50cc IAB WITH ACCESSORIES (TURKEY) D684-00-0296-09 MEGA 8Fr. 50cc IAB WITH ACCESSORIES WITH APA (TURKEY) D684-00-0296-10 MEGA 8Fr. 50cc IAB WITH ACCESSORIES, US ONLY D684-00-0296-01U MEGA 8Fr. 50cc IAB WITH ACCESSORIES, (APA) US ONLY D684-00-0296-02U MEGA 8Fr. 50cc IAB WITH ACCESSORIES (CHINA) D684-00-0498-01 MEGA 8Fr. 50cc IAB WITH ACCESSORIES (INDIA) D684-00-0498-07 MEGA 8Fr. 50cc IAB WITH ACCESSORIES (APA) (INDIA) D684-00-0498-08 PACKAGED INSERTION KIT - MEGA 8 Fr. 50 cc IAB D884-00-0019-17

Reason for recall

During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.

Distribution pattern

Domestic distribution nationwide. International distribution worldwide.

Code information

Product Description UDI MEGA 8Fr. 50cc IAB WITH STATLOCK 10607567108001 MEGA 8Fr. 50cc IAB WITH STATLOCK & APA 10607567108018 MEGA 8Fr. 50cc IAB WITH ACCESSORIES (W/O STATLOCK) 10607567109411 MEGA 8Fr. 50cc IAB WITH ACCESSORIES (TURKEY) 10607567113265 MEGA 8Fr. 50cc IAB WITH ACCESSORIES WITH APA (TURKEY) 10607567113272 MEGA 8Fr. 50cc IAB WITH ACCESSORIES, US ONLY 10607567109619 MEGA 8Fr. 50cc IAB WITH ACCESSORIES, (APA) US ONLY 10607567109626 MEGA 8Fr. 50cc IAB WITH ACCESSORIES (CHINA) 10607567107301 MEGA 8Fr. 50cc IAB WITH ACCESSORIES (INDIA) 10607567109756 MEGA 8Fr. 50cc IAB WITH ACCESSORIES (APA) (INDIA) 10607567109763 PACKAGED INSERTION KIT - MEGA 8 Fr. 50 cc IAB 10607567107493

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2220-2023
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2220-2023. Report a data issue: hello@regdatalab.com.