Z-2222-2020 Class II Terminated

Recalled by Zimmer Biomet, Inc. — Warsaw, IN

Recall Details

Product Type
Devices
Report Date
June 10, 2020
Initiation Date
May 5, 2020
Termination Date
July 15, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3235 total

Product Description

Vital MIS Spinal Fixation system implants CANN POLY EXT TAB 6.5X55MM, Item Number 810M6555 - Product Usage: are non-cervical spinal fixation devices intended for posterior pedicle screw fixation, posterior hook fixation, or anterolateral fixation. Pedicle screw fixation is indicated for skeletally mature patients and for adolescence patients.

Reason for Recall

There are reports of inadequate locking of the extended tab screws. Complaints were reported where the closure top backed out of the screw post-operatively.

Distribution Pattern

US nationwide distribution.

Code Information

UDI: (01)00887868324888; Lot Numbers: C19I0001 C19I0010 C19J0020 C19L0003 C20A0018 SBM108412 SBM118994