Z-2224-2020 Class II Terminated

Recalled by Zimmer Biomet, Inc. — Warsaw, IN

Recall Details

Product Type
Devices
Report Date
June 10, 2020
Initiation Date
May 5, 2020
Termination Date
July 15, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3235 total

Product Description

Vital MIS Spinal Fixation system implants CANN POLY EXT TAB 6.5X40MM, Item Number 810M6540 - Product Usage: are non-cervical spinal fixation devices intended for posterior pedicle screw fixation, posterior hook fixation, or anterolateral fixation. Pedicle screw fixation is indicated for skeletally mature patients and for adolescence patients.

Reason for Recall

There are reports of inadequate locking of the extended tab screws. Complaints were reported where the closure top backed out of the screw post-operatively.

Distribution Pattern

US nationwide distribution.

Code Information

UDI: (01)00887868324857; Lot Numbers: C19H0014 C19I0015 C19J0013 C19K0011 C19L0012 C19L0015 C19L0016 SBM108409 SBM118991 SBM122159