Z-2224-2021 Class II Terminated

Recalled by Integra LifeSciences Corp. — Princeton, NJ

FDA device recall Z-2224-2021 was initiated by Integra LifeSciences Corp. on Jun 1, 2021 and is designated Class II. Reason for recall: Packaging-pouches with wrinkles across the Tyvek side may compromise the integrity of the pouch's sterile barrier The recall status is terminated (terminated Jan 27, 2023). Affected quantity: 906 units.

Recall details

Product Type
Devices
Report Date
Aug 18, 2021
Initiation Date
Jun 1, 2021
Termination Date
Jan 27, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
906 units

Product description

CODMAN RANEY SCALP CLIPS-Disposable cylindrical clips designed to provide hemostasis to the scalp wound edge. Part Number: 201037

Reason for recall

Packaging-pouches with wrinkles across the Tyvek side may compromise the integrity of the pouch's sterile barrier

Distribution pattern

US Nationwide Distribution - Foreign Distribution: Hong Kong

Code information

Lot Numbers: W2102045, W2102046, W2102048 and W2102049 GTIN# 10381780511625

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2224-2021
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2224-2021. Report a data issue: hello@regdatalab.com.