Z-2227-2018 Class II Terminated

Recalled by The Seaberg Company Inc — Wilsonville, OR

FDA device recall Z-2227-2018 was initiated by The Seaberg Company Inc on May 2, 2018 and is designated Class II. Reason for recall: Based on internal testing, results indicated a possible failure of the stitches securing the buckle to the nylon belt could occur, posing a potential risk when used on a human patient to stop arterial… The recall status is terminated (terminated Jul 1, 2020).

Recall details

Product Type
Devices
Report Date
Jun 20, 2018
Initiation Date
May 2, 2018
Termination Date
Jul 1, 2020
Voluntary/Mandated
Voluntary: Firm initiated

Product description

Part #: SAM XT-C, Model: Hi-Viz Orange or Civilian

Reason for recall

Based on internal testing, results indicated a possible failure of the stitches securing the buckle to the nylon belt could occur, posing a potential risk when used on a human patient to stop arterial blood flow. There is a remote chance that a serious injury and/or death could occur as a result of the failure of the device.

Distribution pattern

USA: CA, MD, OH, FL, ID, CO, CT, OR, AZ, IL, ME, MO, NY, SC, MI, AR, MT, CT, WA, TX, KY, NC, NJ, PA OUS: Australia, Canada, China, Columbia, Costa Rica, Europe (EU), Hong Kong, Iceland, India, Iraq, Israel, Japan, Kenya, Korea, Lebanon, Malaysia, Russia, Saudi Arabia, Singapore, South Africa, Thailand, UAE

Code information

Lot #: XT1711, XT1712, XT1715, XT1716, XT1717, XT1719, XT1722, XT1723, XT1724, XT1726, XT1730, XT1731, XT1732, XT1734, XT1735, XT1736, XT1737, XT1738, XT1741, XT1744, XT1745, XT1748, XT1750, XT1751, XT1803, XT1804, XT1806, XT1807, XT1808

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2227-2018
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for The Seaberg Company Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for The Seaberg Company Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about The Seaberg Company Inc →

Source: openFDA record for Z-2227-2018. Report a data issue: hello@regdatalab.com.