Z-2228-2017 Class II Terminated

Recalled by ConvaTec, Inc — Greensboro, NC

FDA device recall Z-2228-2017 was initiated by ConvaTec, Inc on May 10, 2017 and is designated Class II. Reason for recall: Reports of tubing detaching from oxygen masks either prior to use (in packaging) or during use were made by different end users and for different batches. The failures were caused by an improper… The recall status is terminated (terminated Aug 27, 2021). Affected quantity: 816,709.

Recall details

Product Type
Devices
Report Date
Jun 14, 2017
Initiation Date
May 10, 2017
Termination Date
Aug 27, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
816,709

Product description

Oxygen Masks, Elongated (under chin), Adult Med Conc., No Tubing, Product Code 104-E;

Reason for recall

Reports of tubing detaching from oxygen masks either prior to use (in packaging) or during use were made by different end users and for different batches. The failures were caused by an improper solvent bond between the oxygen tubing and the oxygen mask connector. While the defect affects less than 2% of recalled product, there is a potential for harm when oxygen supply to the user is disrupted.

Distribution pattern

CA, CT, FL, GA, IL, MO, NJ, NY, NC, OH, PA, RI, VA , Puerto Rico, Australia, Belgium, Canada, Columbia, Czech, Denmark, Finland, France, Germany, Greece, Israel, Italy, Japan, Jordan, Malaysia, Malta, Netherlands, Norway, Rwanda, Saudi Arabia, Sweden, Switzerland, United Kingdom, and Vietnam

Code information

Lot Number 102153, 102339, 102941, 103176, 103453, 103718, 103919, 104471, 105034, 105282, 105427, 105567, 105996, 106960, 108658, 112024, 112530

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2228-2017
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for ConvaTec, Inc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2228-2017. Report a data issue: hello@regdatalab.com.