Z-2228-2017 Class II Terminated

Recalled by ConvaTec, Inc — Greensboro, NC

Recall Details

Product Type
Devices
Report Date
June 14, 2017
Initiation Date
May 10, 2017
Termination Date
August 27, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
816,709

Product Description

Oxygen Masks, Elongated (under chin), Adult Med Conc., No Tubing, Product Code 104-E;

Reason for Recall

Reports of tubing detaching from oxygen masks either prior to use (in packaging) or during use were made by different end users and for different batches. The failures were caused by an improper solvent bond between the oxygen tubing and the oxygen mask connector. While the defect affects less than 2% of recalled product, there is a potential for harm when oxygen supply to the user is disrupted.

Distribution Pattern

CA, CT, FL, GA, IL, MO, NJ, NY, NC, OH, PA, RI, VA , Puerto Rico, Australia, Belgium, Canada, Columbia, Czech, Denmark, Finland, France, Germany, Greece, Israel, Italy, Japan, Jordan, Malaysia, Malta, Netherlands, Norway, Rwanda, Saudi Arabia, Sweden, Switzerland, United Kingdom, and Vietnam

Code Information

Lot Number 102153, 102339, 102941, 103176, 103453, 103718, 103919, 104471, 105034, 105282, 105427, 105567, 105996, 106960, 108658, 112024, 112530