Z-2230-2023 Class II Ongoing

Recalled by Boston Scientific Corporation — Marlborough, MA

FDA device recall Z-2230-2023 was initiated by Boston Scientific Corporation on May 25, 2023 and is designated Class II. Reason for recall: Manufactured with the incorrect component may result in lower power output and degradation or dimming of the aiming beam, leading to reduced efficiency of fiber performance. Continued use may cause… The recall status is ongoing. Affected quantity: 1245 units.

Recall details

Product Type
Devices
Report Date
Aug 2, 2023
Initiation Date
May 25, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1245 units

Product description

Flexiva Pulse 242 TracTip Single-Use Laser Fibers -intended to be used as a device that transmits Ho:YAG laser energy from cleared laser consoles to urological anatomy. 5-pack UPN: M006L8405961

Reason for recall

Manufactured with the incorrect component may result in lower power output and degradation or dimming of the aiming beam, leading to reduced efficiency of fiber performance. Continued use may cause overheating of the fiber connector, which could result in a burn, if touched

Distribution pattern

Worldwide distribution - US Nationwide and the country of Canada.

Code information

GTIN: 08714729978664 Lot Numbers: 2030305543, 30305544, 30305545, 30305546, 30305547, 30305548, 30305549, 30389370, 30389371, 30389372, 30389373, 30389375, 30389376, 30389377, 30389378, 30429835, 30429837, 30479586, 30479587, 30479589, 30479590, 30521464, 30521466, 30521467, 30521468, 30554456, 30554457. Exp. Date: 7-Oct-2025 to 16-Nov-2025

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2230-2023
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Boston Scientific Corporation? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Boston Scientific Corporation: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Boston Scientific Corporation →

Source: openFDA record for Z-2230-2023. Report a data issue: hello@regdatalab.com.